Suboptimal Handling of Piccolo Samples or Reagent Discs for Consideration in Ebola Response
-
June 17 2019
Source: Emerg Infect Dis. 25(6):1238-1240
Details:
-
Alternative Title:Emerg Infect Dis
-
Personal Author:
-
Description:Operating clinical analyzers within recommended parameters can be challenging during outbreak response. Using the Piccolo Xpress point-of-care blood chemistry analyzer on guinea pig blood, we found that values of many analytes are still readily comparable when samples and reagent discs are handled at various conditions outside of manufacturer recommendations.
-
Subjects:
-
Pubmed ID:30900978
-
Pubmed Central ID:PMC6537746
-
Document Type:
-
Collection(s):
-
Main Document Checksum:
-
Download URL:
-
File Type: