CDC STACKS serves as an archival repository of CDC-published products including scientific findings, journal articles, guidelines, recommendations, or other public health information authored or co-authored by CDC or funded partners.
As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.
i
Suboptimal Handling of Piccolo Samples or Reagent Discs for Consideration in Ebola Response
-
June 17 2019
-
-
Source: Emerg Infect Dis. 25(6):1238-1240
Details:
-
Alternative Title:Emerg Infect Dis
-
Personal Author:
-
Description:Operating clinical analyzers within recommended parameters can be challenging during outbreak response. Using the Piccolo Xpress point-of-care blood chemistry analyzer on guinea pig blood, we found that values of many analytes are still readily comparable when samples and reagent discs are handled at various conditions outside of manufacturer recommendations.
-