Measurements for 8 Common Analytes in Native Sera Identify Inadequate Standardization among 6 Routine Laboratory Assays
Advanced Search
Select up to three search categories and corresponding keywords using the fields to the right. Refer to the Help section for more detailed instructions.

Search our Collections & Repository

All these words:

For very narrow results

This exact word or phrase:

When looking for a specific result

Any of these words:

Best used for discovery & interchangable words

None of these words:

Recommended to be used in conjunction with other fields

Language:

Dates

Publication Date Range:

to

Document Data

Title:

Document Type:

Library

Collection:

Series:

People

Author:

Help
Clear All

Query Builder

Query box

Help
Clear All

For additional assistance using the Custom Query please check out our Help Page

i

Measurements for 8 Common Analytes in Native Sera Identify Inadequate Standardization among 6 Routine Laboratory Assays

Filetype[PDF-645.07 KB]


  • English

  • Details:

    • Alternative Title:
      Clin Chem
    • Description:
      BACKGROUND

      External quality assessment (EQA) with commutable samples is essential for assessing the quality of assays performed by laboratories, particularly when the emphasis is on their standardization status and interchangeability of results.

      METHODS

      We used a panel of 20 fresh-frozen single-donation serum samples to assess assays for the measurement of creatinine, glucose, phosphate, uric acid, total cholesterol, HDL cholesterol, LDL cholesterol, and triglycerides. The commercial random access platforms included: Abbott Architect, Beckman Coulter AU, Ortho Vitros, Roche Cobas, Siemens Advia, and Thermo Scientific Konelab. The assessment was done at the peer group level and by comparison against the all-method trimmed mean or reference method values, where available. The considered quality indicators were intraassay imprecision, combined imprecision (including sample–matrix interference), bias, and total error. Fail/pass decisions were based on limits reflecting state-of-the-art performance, but also limits related to biological variation.

      RESULTS

      Most assays showed excellent peer performance attributes, except for HDL- and LDL cholesterol. Cases in which individual assays had biases exceeding the used limits were the Siemens Advia creatinine (−4.2%), Ortho Vitros phosphate (8.9%), Beckman Coulter AU triglycerides (5.4%), and Thermo Scientific Konelab uric acid (6.4%), which lead to considerable interassay discrepancies. Additionally, large laboratory effects were observed that caused interlaboratory differences of >30%.

      CONCLUSIONS

      The design of the EQA study was well suited for monitoring different quality attributes of assays performed in daily laboratory practice. There is a need for improvement, even for simple clinical chemistry analytes. In particular, the interchangeability of results remains jeopardized both by assay standardization issues and individual laboratory effects.