Reproducibility and Validity of a Handheld Spirometer
Public Domain
-
2008/04/01
Details
-
Personal Author:
-
Description:BACKGROUND: Handheld spirometers have several advantages over desktop spirometers, but worries persist regarding reproducibility and validity of data from handheld spirometers. We undertook an independent examination of the EasyOne handheld spirometer. METHODS: The laboratory testing included reproducibility and validity testing with a waveform generator. We used standard American Thoracic Society waveforms for in-line testing, calibration adaptor testing, and testing during compression of the mouthpiece. The clinical testing involved repeated tests with 24 spirometry-naive volunteers and comparison to spirometry results from laboratory (volume-sensing dry rolling seal) spirometer. RESULTS: The EasyOne exceeded standard thresholds for acceptability with the American Thoracic Society waveforms. In-line testing yielded valid results from the EasyOne. Between the EasyOne and the reference spirometer readings the mean +/- SD difference was 0.03 +/- 0.23 L for forced vital capacity (FVC) and -0.06 +/- 0.09 L for forced expiratory volume in the first second (FEV1). The calibration adaptor showed no appreciable problems. Extreme compression of the mouthpiece reduced the measured values. In clinical testing the coefficients of variation and limits of agreement were, respectively, 3.3% and 0.24 L for FVC, 2.6% and 0.18 L for FEV1, and 1.9% and 0.05 for the FEV1/FVC ratio. The EasyOne readings were lower than those from the reference spirometer; the differences were: -0.12 L for FVC, -0.17 L for FEV1, and -0.02 for FEV1/FVC. The limits of agreement were within criteria for FVC but not for the FEV1, possibly due to a training effect. CONCLUSION: The EasyOne spirometer yielded generally reproducible results that were generally valid, compared to the values from the laboratory spirometer. The use of the EasyOne in clinical, occupational, and research settings seems justified. [Description provided by NIOSH]
-
Subjects:
-
Keywords:
-
ISSN:0020-1324
-
Document Type:
-
Genre:
-
Place as Subject:
-
CIO:
-
Division:
-
Topic:
-
Location:
-
Pages in Document:433-441
-
Volume:53
-
Issue:4
-
NIOSHTIC Number:nn:20033716
-
Citation:Resp Care 2008 Apr; 53(4):433-441
-
Contact Point Address:R. Graham Barr, Department of Medicine, PH-9 East Room 105, Columbia University Medical Center, 622 W 168th Street, New York NY 10032
-
Email:rgb9@columbia.edu
-
Federal Fiscal Year:2008
-
Peer Reviewed:True
-
Source Full Name:Respiratory Care
-
Collection(s):
-
Main Document Checksum:urn:sha-512:4c66c666fbdb7769c0fdff0864ff051beb9b478b46506714dcf8e427d3bdad0bb2e0c752903dea95ac71f3035e2fba60da5da7be61ba99fb1033019f777b9607
-
Download URL:
-
File Type:
ON THIS PAGE
CDC STACKS serves as an archival repository of CDC-published products including
scientific findings,
journal articles, guidelines, recommendations, or other public health information authored or
co-authored by CDC or funded partners.
As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.
As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.
You May Also Like