i
Development and validation of a high-throughput online solid phase extraction – Liquid chromatography – Tandem mass spectrometry method for the detection of tetrodotoxin in human urine
-
May 19 2016
Source: Toxicon. 119:64-71
Details:
-
Alternative Title:Toxicon
-
Personal Author:
-
Description:Tetrodotoxin (TTX) is an extremely potent paralytic toxin responsible for yearly illness and death around the world. A clinical measurement is necessary to confirm exposure because symptoms of TTX intoxication cannot be distinguished from other paralytic toxins. Our group has developed an online solid phase extraction hydrophilic interaction liquid chromatography (HILIC) method for the analysis of TTX in human urine with tandem mass spectrometry. The reportable range for the method was 2.80 - 249 ng/mL in urine with precision and accuracy within 15% as determined for all quality control samples. No isotopically-labeled internal standard is available for TTX; thus a surrogate internal standard, voglibose, was investigated to compensate for matrix effects and ionization suppression. However, upon evaluation, voglibose was ineffective for this purpose. This new online method rapidly identifies TTX, facilitating the work of public health authorities and providing support to monitoring programs worldwide.
-
Subjects:
-
Source:
-
Pubmed ID:27212629
-
Pubmed Central ID:PMC6943750
-
Document Type:
-
Collection(s):
-
Main Document Checksum:
-
Download URL:
-
File Type:
Supporting Files
-
gif jpeg gif jpeg xml gif jpeg gif jpeg gif jpeg gif jpeg