Evaluation of the Alphasense Optical Particle Counter (OPC-N2) and the Grimm Portable Aerosol Spectrometer (PAS-1.108)

Filetype[PDF-736.72 KB]


English

Details:

  • Alternative Title:
    Aerosol Sci Technol
  • Personal Author:
  • Description:
    We compared the performance of a low-cost (∼$500), compact optical particle counter (OPC, OPC-N2, Alphasense) to another OPC (PAS-1.108, Grimm Technologies) and reference instruments. We measured the detection efficiency of the OPCs by size from 0.5 to 5 μm for monodispersed, polystyrene latex (PSL) spheres. We then compared number and mass concentrations measured with the OPCs to those measured with reference instruments for three aerosols: salt, welding fume and Arizona road dust. The OPC-N2 detection efficiency for monodispersed was similar to the PAS-1.108 for particles larger than 0.8 μm (minimum of 79% at 1 μm and maximum of 101% at 3 μm). For 0.5-μm particles, the detection efficiency of OPCN2 was underestimated at 78%, whereas PAS-1.108 overestimated concentrations by 183%. The mass concentrations from the OPCs were linear (r ≥ 0.97) with those from the reference instruments for all aerosols, although the slope and intercept were different. The mass concentrations were overestimated for dust (OPC-N2, slope = 1.6; PAS-1.108, slope = 2.7) and underestimated for welding fume (OPC-N2, slope = 0.05; PAS-1.108, slope = 0.4). The coefficient of variation (CV, precision) for OPC-N2 for all experiments was between 4.2% and 16%. These findings suggest that, given site-specific calibrations, the OPC-N2 can provide number and mass concentrations similar to the PAS-1.108 for particles larger than 1 μm.
  • Source:
  • Pubmed ID:
    28871213