Assessing the Accuracy of Commercially Available Gas Sensors for the Measurement of Ambient Ozone and Nitrogen Dioxide
-
2018/11/01
-
Details
-
Personal Author:
-
Description:The objective of the National Institute for Occupational Safety and Health (NIOSH) accuracy criterion is to ensure that measurements from monitoring devices are within +/- 25% of the true concentration of the analyte with 95% certainty. To determine whether NO2 and O3 sensors meet this criterion, three commercially available units (Cairclip O3/NO2, Aeroqual NO2, and Aeroqual O3 sensors) were co-located three times with validated instruments (NOx chemiluminescence [NO2mon] and photometric O3 analyzers [O3mon]) at an outdoor monitoring station. As cofactors of sensor performance such as temperature (T) and relative humidity (RH) potentially influence the response of NO2 and O3 sensors, corrections for cofactors were made by using T, RH, and the sensor measurements to predict measurements made by NO2mon and O3mon during the first co-location period (training dataset). The developed models were tested in the merged data obtained from the second and third co-location periods (testing dataset). In the training and testing datasets, the mean NO2 as measured by NO2mon was 4.6 ppb (range = 0.4 - 35 ppb) and 9.4 ppb (range = 1 - 37 ppb), respectively. The mean O3 in the training and testing datasets as measured by O3mon was 38.8 ppb (range = 1 - 65 ppb) and 35.7 ppb (range = 1 - 61 ppb), respectively. None of the sensor measurements in the training dataset were within the NIOSH accuracy criterion (mean error >/= 25%). After correcting for cofactors of sensor performance, the accuracy of the Cairclip O3/NO2 and the Aeroqual O3 sensors considerably improved when tested with the testing dataset (mean error = -1% and 14%, respectively). However, the Aeroqual NO2 sensor had an error that was not within +/- 25%. Raw measurements from the tested sensors may be unsuitable for assessing workers' exposure to NO2 and O3. Corrections for cofactors of Cairclip O3/NO2 and Aeroqual O3 sensor performance are required for more accurate occupational exposure assessment. [Description provided by NIOSH]
-
Subjects:
-
Keywords:
-
ISSN:1545-9624
-
Document Type:
-
Funding:
-
Genre:
-
Place as Subject:
-
CIO:
-
Topic:
-
Location:
-
Volume:15
-
Issue:11
-
NIOSHTIC Number:nn:20052616
-
Citation:J Occup Environ Hyg 2018 Nov; 15(11):782-791
-
Contact Point Address:Tiina Reponen, Department of Environmental Health, University of Cincinnati, Cincinnati, Ohio, USA
-
Email:reponeta@ucmail.uc.edu
-
CAS Registry Number:
-
Federal Fiscal Year:2019
-
Performing Organization:University of Cincinnati
-
Peer Reviewed:True
-
Start Date:20050701
-
Source Full Name:Journal of Occupational and Environmental Hygiene
-
End Date:20260630
-
Collection(s):
-
Main Document Checksum:urn:sha-512:63dabcc1a54103d1756061033dee7c9e1cc2684f1cd41281cd61de0785fc2a4248a0f35b83759a3003528f772cd125d40117e0c04094080eff5fb436761ff3ed
-
Download URL:
-
File Type:
ON THIS PAGE
CDC STACKS serves as an archival repository of CDC-published products including
scientific findings,
journal articles, guidelines, recommendations, or other public health information authored or
co-authored by CDC or funded partners.
As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.
As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.
You May Also Like