Evaluation of Low-Cost Electro-Chemical Sensors for Environmental Monitoring of Ozone, Nitrogen Dioxide, and Carbon Monoxide
-
2018/02/01
-
Details
-
Personal Author:
-
Description:Development of an air quality monitoring network with high spatio-temporal resolution requires installation of a large number of air pollutant monitors. However, state-of-the-art monitors are costly and may not be compatible with wireless data logging systems. In this study, low-cost electro-chemical sensors manufactured by Alphasense Ltd. for detection of CO and oxidative gases (predominantly O3 and NO2) were evaluated. The voltages from three oxidative gas sensors and three CO sensors were recorded every 2.5 s when exposed to controlled gas concentrations in a 0.125-m3 acrylic glass chamber. Electro-chemical sensors for detection of oxidative gases demonstrated sensitivity to both NO2 and O3 with similar voltages recorded when exposed to equivalent environmental concentrations of NO2 or O3 gases, when evaluated separately. There was a strong linear relationship between the recorded voltages and target concentrations of oxidative gases (R2 > 0.98) over a wide range of concentrations. Although a strong linear relationship was also observed for CO concentrations below 12 ppm, a saturation effect was observed wherein the voltage only changes minimally for higher CO concentrations (12 to 50 ppm). The non-linear behavior of the CO sensors implied their unsuitability for environments where high CO concentrations are expected. Using a manufacturer-supplied shroud, sensors were tested at 2 different flow rates (0.25 and 0.5 Lpm) to mimic field calibration of the sensors with zero air and a span gas concentration (2 ppm NO2 or 15 ppm CO). As with all electrochemical sensors, the tested devices were subject to drift with a bias up to 20% after 9 months of continuous operation. Alphasense CO sensors were found to be a proper choice for occupational and environmental CO monitoring with maximum concentration of 12 ppm, especially due to the field-ready calibration capability. Alphasense oxidative gas sensors are usable only if it is valuable to know the sum of the NO2 and O3 concentrations. [Description provided by NIOSH]
-
Subjects:
-
Keywords:
-
ISSN:1545-9624
-
Document Type:
-
Funding:
-
Genre:
-
Place as Subject:
-
CIO:
-
Topic:
-
Location:
-
Pages in Document:87-98
-
Volume:15
-
Issue:2
-
NIOSHTIC Number:nn:20050565
-
Citation:J Occup Environ Hyg 2018 Feb; 15(2):87-98
-
Contact Point Address:Kirsten Koehler, PhD, Department of Environmental Health and Engineering, Johns Hopkins Bloomberg School of Public Health, 615 N. Wolfe St., Room E6632, Baltimore, MD 21205
-
Email:kirsten.koehler@jhu.edu
-
CAS Registry Number:
-
Federal Fiscal Year:2018
-
NORA Priority Area:
-
Performing Organization:Johns Hopkins University, Baltimore, Maryland
-
Peer Reviewed:True
-
Start Date:20140901
-
Source Full Name:Journal of Occupational and Environmental Hygiene
-
End Date:20180831
-
Collection(s):
-
Main Document Checksum:urn:sha-512:15e760885e94eace7cf6bc307df42f26f829a42885ea9e5b4fff743c6d3e8f04fb6383e012a25954945fbc9605f6fcb652a5db9e5186507348ab525bb7c402fe
-
Download URL:
-
File Type:
ON THIS PAGE
CDC STACKS serves as an archival repository of CDC-published products including
scientific findings,
journal articles, guidelines, recommendations, or other public health information authored or
co-authored by CDC or funded partners.
As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.
As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.
You May Also Like