Evaluation of pump pulsation in respirable size-selective sampling: part I. Pulsation measurements

Public Domain


Details

  • Personal Author:
  • Description:
    Pulsations generated by personal sampling pumps modulate the airflow through the sampling trains, thereby varying sampling efficiencies, and possibly invalidating collection or monitoring. The purpose of this study was to characterize pulsations generated by personal sampling pumps relative to a nominal flow rate at the inlet of different respirable cyclones. Experiments were conducted using a factorial combination of 13 widely used sampling pumps (11 medium and 2 high volumetric flow rate pumps having a diaphragm mechanism) and 7 cyclones [10-mm nylon also known as Dorr- Oliver (DO), Higgins-Dewell (HD), GS-1, GS-3, Aluminum, GK2.69, and FSP-10]. A hot-wire anemometer probe cemented to the inlet of each cyclone type was used to obtain pulsation readings. The three medium flow rate pump models showing the highest, a midrange, and the lowest pulsations and two high flow rate pump models for each cyclone type were tested with dust-loaded filters (0.05, 0.21, and 1.25 mg) to determine the effects of filter loading on pulsations. The effects of different tubing materials and lengths on pulsations were also investigated. The fundamental frequency range was 22.110 Hz and the magnitude of pulsation as a proportion of the mean flow rate ranged from 4.4 to 73.1%. Most pump/cyclone combinations generated pulse magnitudes .10% (48 out of 59 combinations), while pulse shapes varied considerably. Pulsation magnitudes were not considerably different for the clean and dust-loaded filters for the DO, HD, and Aluminum cyclones, but no consistent pattern was observed for the other cyclone types. Tubing material had less effect on pulsations than tubing length; when the tubing length was 183 cm, pronounced damping was observed for a pump with high pulsation (>60%) for all tested tubing materials except for the Tygon Inert tubing. The findings in this study prompted a further study to determine the possibility of shifts in cyclone sampling efficiency due to sampling pump pulsations, and those results are reported subsequently. [Description provided by NIOSH]
  • Subjects:
  • Keywords:
  • ISSN:
    0003-4878
  • Document Type:
  • Genre:
  • Place as Subject:
  • CIO:
  • Division:
  • Topic:
  • Location:
  • Pages in Document:
    60-73
  • Volume:
    58
  • Issue:
    1
  • NIOSHTIC Number:
    nn:20043210
  • Citation:
    Ann Occup Hyg 2014 Jan; 58(1):60-73
  • Contact Point Address:
    Eun Gyung Lee, National Institute for Occupational Safety and Health, Health Effects Laboratory Division, Exposure Assessment Branch, 1095 Willowdale Road, Morgantown, WV 26505
  • Email:
    dtq5@cdc.gov
  • Federal Fiscal Year:
    2014
  • NORA Priority Area:
  • Peer Reviewed:
    True
  • Source Full Name:
    Annals of Occupational Hygiene
  • Collection(s):
  • Main Document Checksum:
    urn:sha-512:ddf07662ecd59de2f51c6ef55e7f863ea32bbaa9b20334f9d742644ff695e6123fdbaa613e1f6e6b83ab9221e23f733975e9fb2c30c99ced4b5c352ea540f994
  • Download URL:
  • File Type:
    Filetype[PDF - 3.31 MB ]
ON THIS PAGE

CDC STACKS serves as an archival repository of CDC-published products including scientific findings, journal articles, guidelines, recommendations, or other public health information authored or co-authored by CDC or funded partners.

As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.