Reliability of a widely used test of peripheral cutaneous vibration sensitivity and a comparison of two testing protocols.



Details

  • Personal Author:
  • Description:
    Tests were carried out to investigate the reliability and time efficiency of two testing protocols for the determination of cutaneous vibration sensitivity. Twenty two healthy volunteers were tested using the Vibratron-II and two psychophysical procedures, the forced choice (SFC) and the method of limits (MOL). Both methods were performed on the same subjects on two occasions to compare their reliability and time efficiency. None of the subjects had a history of serious medical illness. Vibration thresholds determined by the SFC procedure were substantially lower than those determined by the MOL. The correlation of vibration thresholds between the two sessions, an index of the method's reliability, was much higher for the MOL than the SFC procedure. The SFC procedure took about twice as long to administer as the MOL procedure. The average vibration threshold was slightly lower in the second MOL session when compared with the first, suggesting some systematic criterion shift between occasions, perhaps related to learning about how to respond. The difference in mean vibration threshold between sessions for the SFC procedure was negligible. The authors conclude that the current widespread uncritical acceptance of forced choice procedures as preferable to the method of limits may be unwarranted. The authors recommend a method of limits procedure when the severity of the disease in the population being tested is mild to moderate and when the subjects are able to understand and comply with the instructions. [Description provided by NIOSH]
  • Subjects:
  • Keywords:
  • ISSN:
    0007-1072
  • Document Type:
  • Funding:
  • Genre:
  • Place as Subject:
  • CIO:
  • Topic:
  • Location:
  • Volume:
    45
  • Issue:
    9
  • NIOSHTIC Number:
    nn:00183564
  • Citation:
    Br J Ind Med 1988 Sep; 45(9):635-639
  • Contact Point Address:
    Community Medicine Mount Sinai School of Medicine One Gustave L Levy Place New York, N Y 10029
  • Federal Fiscal Year:
    1988
  • NORA Priority Area:
  • Performing Organization:
    Mount Sinai School of Medicine, New York, New York
  • Peer Reviewed:
    True
  • Start Date:
    19861101
  • Source Full Name:
    British Journal of Industrial Medicine
  • End Date:
    19870831
  • Collection(s):
  • Main Document Checksum:
    urn:sha-512:fc22973f1191e89385b8386d6997fee23fa8d4af671e1da9f1c8fe8ab07a9a37265885a2cdf0000e2fb2b4a86194264b3f5ae215e622403f4ed200deab158fec
  • Download URL:
  • File Type:
    Filetype[PDF - 822.80 KB ]
ON THIS PAGE

CDC STACKS serves as an archival repository of CDC-published products including scientific findings, journal articles, guidelines, recommendations, or other public health information authored or co-authored by CDC or funded partners.

As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.