Quality control of multiplex antibody detection in samples from large-scale surveys: the example of malaria in Haiti
Supporting Files
Public Domain
-
24 2020
-
File Language:
English
Details
-
Alternative Title:Sci Rep
-
Personal Author:van den Hoogen, Lotus L. ; Présumé, Jacquelin ; Romilus, Ithamare ; Mondélus, Gina ; Elismé, Tamara ; Sepúlveda, Nuno ; Stresman, Gillian ; Druetz, Thomas ; Ashton, Ruth A. ; Joseph, Vena ; Eisele, Thomas P. ; Hamre, Karen E. S. ; Chang, Michelle A. ; Lemoine, Jean F. ; Tetteh, Kevin K. A. ; Boncy, Jacques ; Existe, Alexandre ; Drakeley, Chris ; Rogier, Eric
-
Description:Measuring antimalarial antibodies can estimate transmission in a population. To compare outputs, standardized laboratory testing is required. Here we describe the in-country establishment and quality control (QC) of a multiplex bead assay (MBA) for three sero-surveys in Haiti. Total IgG data against 21 antigens were collected for 32,758 participants. Titration curves of hyperimmune sera were included on assay plates, assay signals underwent 5-parameter regression, and inspection of the median and interquartile range (IQR) for the y-inflection point was used to determine assay precision. The medians and IQRs were similar for Surveys 1 and 2 for most antigens, while the IQRs increased for some antigens in Survey 3. Levey-Jennings charts for selected antigens provided a pass/fail criterion for each assay plate and, of 387 assay plates, 13 (3.4%) were repeated. Individual samples failed if IgG binding to the generic glutathione-S-transferase protein was observed, with 659 (2.0%) samples failing. An additional 455 (1.4%) observations failed due to low bead numbers (<20/analyte). The final dataset included 609,438 anti-malaria IgG data points from 32,099 participants; 96.6% of all potential data points if no QC failures had occurred. The MBA can be deployed with high-throughput data collection and low inter-plate variability while ensuring data quality.
-
Subjects:
-
Source:Sci Rep. 10
-
Pubmed ID:31980693
-
Pubmed Central ID:PMC6981173
-
Document Type:
-
Volume:10
-
Collection(s):
-
Main Document Checksum:urn:sha256:a40cf1e66bacb9ae686073f98e83e710e165ed3fd6ca5c7de8939a7200209402
-
Download URL:
-
File Type:
Supporting Files
File Language:
English
ON THIS PAGE
CDC STACKS serves as an archival repository of CDC-published products including
scientific findings,
journal articles, guidelines, recommendations, or other public health information authored or
co-authored by CDC or funded partners.
As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.
As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.
You May Also Like
COLLECTION
CDC Public Access