Effects of preanalytical factors on hemoglobin measurement: a comparison of two HemoCue® point-of-care analyzers
-
Apr 12 2017
-
-
Source: Clin Biochem. 50(9):513-520.
Details:
-
Alternative Title:Clin Biochem
-
Personal Author:
-
Description:Background
In field studies, hemoglobin (Hb) is often measured using a battery-operated, portable HemoCue® hemoglobinometer.
Methods
We compared the performance of 2 HemoCue® models (Hb-201+ and Hb-301) and investigated effects of preanalytical factors on Hb results by simulating unfavorable field conditions.
Results
The Hb-301 produced 2.6% higher results compared to the Hb-201+. Hb had to be measured within 1 min of filling the Hb-301 cuvette to avoid artificially elevated concentrations (1.3% per min). The Hb-301 cuvettes withstood elevated temperature (37°C) and humidity (72%) for 3 wk, while the Hb-201+ cuvettes degraded within 10 min under those conditions. Both cuvette types withstood elevated temperature for 3 wk. Properly-collected venous and capillary blood produced comparable results. Pooled capillary blood produced comparable results to the second and third but not the fourth drop of blood (3.3% lower). Blood could be stored for ≤4 d at 10–30°C before Hb-201+ measurement, but only for 1 d at 10–23°C before Hb-301 measurement (≤1% change in Hb).
Conclusions
Higher Hb results obtained with the Hb-301 may influence the interpretation of anemia prevalence in health surveys. While the Hb-301 performed better in high humidity conditions, the Hb-201+ provided more user flexibility regarding delayed Hb reading.
-
Subjects:
-
Source:
-
Pubmed ID:28412284
-
Pubmed Central ID:PMC5685159
-
Document Type:
-
Funding:
-
Collection(s):
-
Main Document Checksum:
-
Download URL:
-
File Type: