Calibration of high flow rate thoracic-size selective samplers

Supporting Files
File Language:
English


Details

  • Alternative Title:
    J Occup Environ Hyg
  • Personal Author:
  • Description:
    High flow rate respirable size selective samplers, GK4.126 and FSP10 cyclones, were calibrated for thoracic-size selective sampling in two different laboratories. The National Institute for Occupational Safety and Health (NIOSH) utilized monodisperse ammonium fluorescein particles and scanning electron microscopy to determine the aerodynamic particle size of the monodisperse aerosol. Fluorescein intensity was measured to determine sampling efficiencies of the cyclones. The Health Safety and Laboratory (HSL) utilized a real time particle sizing instrument (Aerodynamic Particle Sizer) and polydisperse glass sphere particles and particle size distributions between the cyclone and reference sampler were compared. Sampling efficiency of the cyclones were compared to the thoracic convention defined by the American Conference of Governmental Industrial Hygienists (ACGIH)/Comité Européen de Normalisation (CEN)/International Standards Organization (ISO). The GK4.126 cyclone showed minimum bias compared to the thoracic convention at flow rates of 3.5 l min(-1) (NIOSH) and 2.7-3.3 l min(-1) (HSL) and the difference may be from the use of different test systems. In order to collect the most dust and reduce the limit of detection, HSL suggested using the upper end in range (3.3 l min(-1)). A flow rate of 3.4 l min(-1) would be a reasonable compromise, pending confirmation in other laboratories. The FSP10 cyclone showed minimum bias at the flow rate of 4.0 l min(-1) in the NIOSH laboratory test. The high flow rate thoracic-size selective samplers might be used for higher sample mass collection in order to meet analytical limits of quantification.
  • Subjects:
  • Source:
    J Occup Environ Hyg. 13(6):D93-D98.
  • Pubmed ID:
    26891196
  • Pubmed Central ID:
    PMC4890162
  • Document Type:
  • Funding:
  • Place as Subject:
  • Volume:
    13
  • Issue:
    6
  • Collection(s):
  • Main Document Checksum:
    urn:sha256:20ea046d8de2a6244198b0d07e470e0c4e12b560542dab2c8565f04da18bdbe7
  • Download URL:
  • File Type:
    Filetype[PDF - 235.09 KB ]
File Language:
English
ON THIS PAGE

CDC STACKS serves as an archival repository of CDC-published products including scientific findings, journal articles, guidelines, recommendations, or other public health information authored or co-authored by CDC or funded partners.

As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.