Accuracy and Precision of Several Portable Gas Detectors - a Supplement to RI 7811 [RI 8010]
Public Domain
-
1975/01/01
-
Series: Mining Publications
File Language:
English
Details
-
Personal Author:
-
Description:The performance of commercially available portable instrumentation for detecting nitrogen dioxide, sulfur dioxide, and hydrogen sulfide was evaluated. Length-of-stain detector tubes from several manufacturers were included in the study. The instruments were evaluated in atmospheres with 50- and 80-percent relative humidities at 75 deg f and containing test gas concentrations of 0.1, 0.25, 0.50., 1.0, 1.5, and 2.0 Times the threshold limit values. The results from the gas analyzers were correlated with the known contaminant concentration by calculating and plotting an estimated line of regression, confidence limits, and correlation coefficient for each detector system. An increase in the relative humidity from 50 to 80 percent at 75 deg f changed the instrument response to the contaminants. The uncertainties in detection precision reported are for the observer-instrument system. The main difference or inconsistency in toxic gas concentrations determined with length-of- stain tubes resulted from the observer's interpretation of the results rather than from a predominant difference in the tubes. [Description provided by NIOSH]
-
Content Notes:in NTRL, no pdf
-
Subjects:
-
Series:
-
Subseries:
-
Document Type:
-
Genre:
-
CIO:
-
Division:
-
Topic:
-
Location:
-
Pages in Document:1-34
-
NIOSHTIC Number:nn:10000321
-
NTIS Accession Number:PB240848
-
Citation:NTIS: PB 240 848; :1-34
-
Federal Fiscal Year:1975
-
Peer Reviewed:False
-
Collection(s):
-
Main Document Checksum:urn:sha-512:d6f59e1a6f212120f727251c5af7c7784a2627b5dd0df756fc249e8367985241bbba3773ce3d60800f86724513d3e0ea957ef99e28d24898b4c701cc8400ef85
-
Download URL:
-
File Type:
File Language:
English
ON THIS PAGE
CDC STACKS serves as an archival repository of CDC-published products including
scientific findings,
journal articles, guidelines, recommendations, or other public health information authored or
co-authored by CDC or funded partners.
As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.
As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.
You May Also Like