A portable, high-speed, vacuum-outlet GC vapor analyzer employing air as carrier gas and surface acoustic wave detection



Details

  • Personal Author:
  • Description:
    Vacuum-outlet GC with atmospheric-pressure air as the carrier gas is implemented at outlet pressures up to 0.8 atm using a low-dead-volume polymer-coated surface acoustic wave (SAW) detector. Increases in the system outlet pressure from 0.1 to 0.8 atm lead to proportional increases in detector sensitivity and significant increases in column efficiency. The latter effect arises from the fact that optimal carrier gas velocities are lower in air than in more conventional carrier gases such as helium or hydrogen due to the smaller binary diffusion coefficients of vapors in air. A 12-m-long, 0.25-mm-i.d. tandem column ensemble consisting of 4.5-m dimethyl polysiloxane and 7.5-m trifluoropropylmethyl polysiloxane operated at an outlet pressure of 0.5 atm provides up to 4 x 10 4 theoretical plates and a peak capacity of 65 (resolution, 1.5) for a 3-min isothermal analysis. At 30 degrees C, mixtures of vapors ranging in vapor pressure from 8.6 to 76 Torr are separated in this time frame. The SAW detector cell has an internal volume of <2 uL, which allows the use of higher column outlet pressures with minimal dead time. The sensor response is linear with solute mass over at least 2-3 decades and provides detection limits of 20-50 ng for the vapors tested. The combination of atmospheric-pressure air as carrier gas, modest operating pressures, and SAW sensor detection is well-suited for field instrumentation since it eliminates the need for support gases, permits smaller, low-power pumps to be used, and provides sensitivity to a wide range of vapor analytes. [Description provided by NIOSH]
  • Subjects:
  • Keywords:
  • ISSN:
    0003-2700
  • Document Type:
  • Funding:
  • Genre:
  • Place as Subject:
  • CIO:
  • Topic:
  • Location:
  • Volume:
    73
  • Issue:
    19
  • NIOSHTIC Number:
    nn:20042728
  • Citation:
    Anal Chem 2001 Oct; 73(19):4668-4675
  • CAS Registry Number:
  • Federal Fiscal Year:
    2002
  • Performing Organization:
    University of Michigan, Ann Arbor
  • Peer Reviewed:
    True
  • Start Date:
    20060601
  • Source Full Name:
    Analytical Chemistry
  • End Date:
    20060430
  • Collection(s):
  • Main Document Checksum:
    urn:sha-512:fcbdce59c8fa42b6c2ef4896db75ba44e22761b83d12236ab89ecea81049b373fd841800d20282d5638bfb3c12112b0780089e25e9d1c2f6bf7c717f619a3f26
  • Download URL:
  • File Type:
    Filetype[PDF - 100.60 KB ]
ON THIS PAGE

CDC STACKS serves as an archival repository of CDC-published products including scientific findings, journal articles, guidelines, recommendations, or other public health information authored or co-authored by CDC or funded partners.

As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.