Four-Year Calibration Stability of the EasyOne Portable Spirometer



Details

  • Personal Author:
  • Description:
    BACKGROUND: Clinical practice guidelines recommend daily spirometer calibration checks and weekly linearity checks. The long-term stability of the volume and flow accuracy of a specific model of spirometer should be carefully characterized before modification of the frequency of calibration checks is considered for that model of spirometer. METHODS: The EasyOne ultrasonic flow-sensing spirometer was chosen for use by the clinical centers at the 2002 inception of the World Trade Center Worker and Volunteer Medical Screening Program. The screening program quality-control procedure required that the expiratory and inspiratory volume accuracy of each spirometer be checked every day of testing, and the flow accuracy (linearity) checked every week. The calibration check results were transferred to a central database for summary. RESULTS: Over 5,000 calibration-check results (4,109 single-speed and 1,189 three-speed) were accumulated from a total of 34 spirometers during the period February 2003 through March 2007. The mean single-speed calibration errors (and 5th-95th percentiles) were -2 mL (-80 to 70 mL) for exhalation and -10 mL (-80 to 60 mL) for inhalation. 98% of the exhalation and 97% of the inhalation calibration checks were accurate within 3.0%. There was no evidence of significant non-linearity according to the results of the 3-speed calibration checks (mean errors of -3, -5, and -6 mL at each speed). CONCLUSIONS: The EasyOne retained inhalation and exhalation volume accuracy of better than 3% for at least 4 years. Routine multiple-speed volume calibration checks may not be necessary with the EasyOne. The acceptability and repeatability of patient efforts should be the primary focus of quality-assurance programs with spirometers that have been demonstrated to remain accurate for long periods. [Description provided by NIOSH]
  • Subjects:
  • Keywords:
  • ISSN:
    0020-1324
  • Document Type:
  • Funding:
  • Genre:
  • Place as Subject:
  • CIO:
  • Division:
  • Topic:
  • Location:
  • Volume:
    55
  • Issue:
    7
  • NIOSHTIC Number:
    nn:20037258
  • Citation:
    Respir Care 2010 Jul; 55(7):873-877
  • Contact Point Address:
    Paul L Enright, National Institute for Occupational Safety and Health, Centers for Disease Control, Morgantown, West Virginia
  • Email:
    lungguy@aol.com
  • Federal Fiscal Year:
    2010
  • Performing Organization:
    Mount Sinai School of Medicine of New York University
  • Peer Reviewed:
    True
  • Start Date:
    20040601
  • Source Full Name:
    Respiratory Care
  • End Date:
    20100531
  • Collection(s):
  • Main Document Checksum:
    urn:sha-512:83b5057eed16169851cfa600c8615697a28018a0851dea69a496c4653461b17c7ce708a87f12b53306f3348f9a1a8d279ca9ce6f2ac0c73f87c3d37d4b9c55a3
  • Download URL:
  • File Type:
    Filetype[PDF - 201.46 KB ]
ON THIS PAGE

CDC STACKS serves as an archival repository of CDC-published products including scientific findings, journal articles, guidelines, recommendations, or other public health information authored or co-authored by CDC or funded partners.

As a repository, CDC STACKS retains documents in their original published format to ensure public access to scientific information.