i
Evaluation of Surgical N95 Respirators Covered with Combinations of Masks and Face Shield
-
7 2023
-
-
Source: J Occup Environ Med. 65(7):610-614
Details:
-
Alternative Title:J Occup Environ Med
-
Personal Author:
-
Description:OBJECTIVE:
Evaluation of the metabolic demands and internal breathing environments when covering an N95 with a surgical mask (SM), cloth mask (CM), and/or face shield (FS).
METHODS:
Three N95 models approved by the National Institute for Occupational Safety and Health (NIOSH) were evaluated under six covering conditions using the NIOSH Automated Breathing and Metabolic Simulator. All conditions used one trial with each N95 for six incremental 5-minute work rates. Inhaled oxygen and carbon dioxide concentrations, peak inhaled and exhaled pressures, and inhaled wet bulb and dry bulb temperatures were measured continuously and averaged across all work rates and covering conditions.
CONCLUSIONS:
Results suggest that metabolic demands and internal breathing environments are significantly impacted by all combinations of coverings tested when compared to N95 only.
-
Subjects:
-
Source:
-
Pubmed ID:37043393
-
Pubmed Central ID:PMC10330396
-
Document Type:
-
Funding:
-
Volume:65
-
Issue:7
-
Collection(s):
-
Main Document Checksum:
-
Download URL:
-
File Type: