Evaluation of a portable gas chromatograph with photoionization detector under variations of VOC concentration, temperature, and relative humidity
-
Apr 2018
Source: J Occup Environ Hyg. 15(4):351-360 -
Alternative Title:J Occup Environ Hyg
-
Personal Author:
-
Description:The objective of this present study was to evaluate the performance of a portable gas chromatograph-photoionization detector (GC-PID), under various test conditions to determine if it could be used in occupational settings. A mixture of 7 volatile organic compounds (VOCs)-acetone, ethylbenzene, methyl isobutyl ketone, toluene, m-xylene, p-xylene, and o-xylene-was selected because its components are commonly present in paint manufacturing industries. A full-factorial combination of 4 concentration levels (exposure scenarios) of VOC mixtures, 3 different temperatures (25°C, 30°C, and 35°C), and 3 relative humidities (RHs; 25%, 50%, and 75%) was conducted in a full-size controlled environmental chamber. Three repetitions were conducted for each test condition allowing for estimation of accuracy. Time-weighted average exposure data were collected using solid sorbent tubes (Anasorb 747, SKC Inc.) as the reference sampling medium. Calibration curves of Frog-4000 using the dry gases showed R| > 0.99 for all analytes except for toluene (R| = 0.97). Frog-4000 estimates within a test condition showed good consistency for the performance of repeated measurement. However, there was ∼41-64% reduction in the analysis of polar acetone with 75% RH relative to collection at 25% RH. Although Frog-4000 results correlated well with solid sorbent tubes (r = 0.808-0.993, except for toluene) most of the combinations regardless of analyte did not meet the <25% accuracy criterion recommended by NIOSH. The effect of chromatographic co-elution can be seen with m, p-xylene when the results are compared to the sorbent tube sampling technique with GC-flame ionization detector. The results indicated an effect of humidity on the quantification of the polar compounds that might be attributed to the pre-concentrator placed in the selected GC-PID. Further investigation may resolve the humidity effect on sorbent trap with micro GC pre-concentrator when water vapor is present. Although this instrument does not fulfill the accuracy criterion specified in the NIOSH technical report No. 2012-162, it can be used as a screening tool for range finding monitoring with dry gases calibration in the occupational setting rather than compliance monitoring.
-
Subjects:
-
Pubmed ID:29333991
-
Pubmed Central ID:PMC6379918
-
Document Type:
-
Collection(s):
-
Main Document Checksum:
-
Download URL:
-
File Type:
-
gif jpeg gif jpeg gif jpeg gif jpeg gif jpeg gif gif jpeg gif jpeg gif jpeg xml jpeg gif jpeg gif jpeg gif jpeg
Details:
Supporting Files
More +