Ambulances are frequently contaminated with infectious microorganisms shed by patients during transport that can be transferred to subsequent patients and emergency medical service workers. Manual decontamination is tedious and time-consuming, and persistent contamination is common even after cleaning. Ultraviolet germicidal irradiation (UVGI) has been proposed as a terminal disinfection method for ambulance patient compartments. However, no published studies have tested the use of UVGI in ambulances. The objectives of this study were to investigate the efficacy of a UVGI system in an ambulance patient compartment and to examine the impact of UVGI fixture position and the UV reflectivity of interior surfaces on the time required for disinfection. A UVGI fixture was placed in the front, middle or back of an ambulance patient compartment, and the UV irradiance was measured at 49 locations. Aluminum sheets and UV-reflective paint were added to examine the effects of increasing surface reflectivity on disinfection time. Disinfection tests were conducted using
Our results showed that the UV irradiance varied considerably depending upon the surface location. For example, with the UVGI fixture in the back position and without the addition of UV-reflective surfaces, the most irradiated location received a dose of UVGI sufficient for disinfection in 16 seconds, but the least irradiated location required 15 hours. Because the overall time required to disinfect all of the interior surfaces is determined by the time required to disinfect the surfaces receiving the lowest irradiation levels, the patient compartment disinfection times for different UVGI configurations ranged from 16.5 hours to 59 minutes depending upon the UVGI fixture position and the interior surface reflectivity. These results indicate that UVGI systems can reduce microbial surface contamination in ambulance compartments, but the systems must be rigorously validated before deployment. Optimizing the UVGI fixture position and increasing the UV reflectivity of the interior surfaces can substantially improve the performance of a UVGI system and reduce the time required for disinfection.
Ambulance patient compartments are frequently contaminated with pathogenic microorganisms shed by patients during transport. These microorganisms can potentially be transferred to subsequent patients and to emergency medical service (EMS) workers by direct contact with the surfaces or by indirect transmission via hands or medical items. The potential risk for the transmission of infections via contaminated surfaces is of great concern to the EMS community and could become a critical problem during an infectious disease pandemic, when large numbers of highly contagious patients would be transported and when the ability to decontaminate ambulances and return them to service as soon as possible would be needed. In 2012, the InterAgency Board, a working group of emergency preparedness and response officials, listed the development of rapid decontamination systems for ambulances as one of its research priorities.(
Current procedures for infection prevention in emergency medical services typically call for the interior of the patient compartment to be cleaned of visible bodily fluids or soil, sprayed with an approved disinfectant, wiped after the appropriate contact time, and allowed to air dry.(
One potential method to reduce the risk of disease transmission in patient compartments is disinfection through the use of whole-compartment systems, such as disinfectant foggers or ultraviolet germicidal irradiation (UVGI). Such germicidal systems are employed as a final disinfection step after the patient has been removed and visible contamination has been cleaned (called terminal disinfection). Terminal disinfection systems do not eliminate the need to first manually clean heavy or visible contamination. However, the use of a terminal disinfection system could allow the manual cleaning to focus on surfaces that are visibly contaminated or most prone to contamination, rather than attempting to wipe down every surface in the ambulance.
UVGI has been studied as a method for surface disinfection in hospital rooms (reviewed by Weber et al.(
UVGI systems have several potential advantages for terminal disinfection. They are relatively simple and easy to use, and do not leave chemical residues or risk exposing patients and workers to toxic chemicals. In ambulances, UVGI systems can be used while the crew cab is occupied, while fogging systems cannot. However, UVGI systems also have an important limitation: Because most materials are not good reflectors of ultraviolet light at germicidal wavelengths (primarily 254 nm), UVGI systems are less effective against microorganisms on surfaces that are not in a direct line-of-sight of the system.(
Although terminal disinfection systems are increasingly popular for use in patient rooms in hospitals,(
While information about the use of UVGI in hospital rooms is limited, information about the use of UVGI in ambulances is virtually non-existent. Although UVGI systems are being marketed as a means of surface disinfection in ambulances, a search of the biomedical literature located only one study which mentioned that UVGI was used for terminal disinfection of ambulances used for transportation of patients with Ebola virus disease, and in that case no test results or details about the system were provided.(
The purpose of this project was to test the ability of an ultraviolet germicidal irradiation system to disinfect the interior of an ambulance patient compartment, to examine the variations in irradiance among different locations in the compartment, and to study the effects of the location of the UVGI fixture and the addition of UV-reflective material or UV-reflective paint on the efficacy of the UVGI system. The information provided by this study will help to better understand the uses and limitations of UVGI systems in ambulances.
Our study was conducted using a 2005 Wheeled Coach Type III ambulance which met Federal Specification KKK-A-1822D when constructed.(
The ultraviolet germicidal light fixture used in these experiments was custom-built. It consisted of ten UV-C lamps with a primary wavelength of 254 nm (TUV PL-L 60 W/4P HO 1CT/25, Philips Lighting). Each lamp had a nominal wattage of 60 watts and a UV wattage of 12.4 watts. The lamps were mounted vertically in two circles (one upper, one lower) around an 11.1 cm (4 3/8”) diameter aluminum-covered post. The lamps were powered by five electronic ballasts (PureVOLT IUV-2S60-M4-LD, Philips Lighting). Photographs and a diagram of the fixture are included in the
The light fixture was placed in three positions in the patient compartment for testing: Front, with the light 81 cm (32“) from the left interior wall and 217 cm (85 1/2”) forward of the rear doors of ambulance; Middle, with the light centered laterally between the interior walls and 132 cm (52”) from the rear doors; and Back, with the light position centered laterally between the interior walls and 48 cm (19") from the rear doors. The light fixture positions are shown in the ambulance patient compartment schematic in the
Irradiance measurements were made at 49 locations throughout the patient compartment as shown in
The irradiance measurements were made using 17 UV-C sensors (SED033–25, International Light Technologies) routed through three automated switch boxes (A803, International Light Technologies) connected to three radiometers (ILT-1700, International Light Technologies). A scan delay of 3 seconds was used for the switch boxes to prevent measurements from one sensor from being carried over to the next sensor reading.
To perform irradiance measurements, the sensors were placed at 17 locations. The UVGI light fixture was turned on and allowed to stabilize for 30 minutes. Separate irradiance measurements were collected with the fixture in the back, middle and front positions. The fixture was then turned off and 16 of the sensors were moved to different locations; one sensor was left in the same location (B2) for all tests to allow comparison of irradiance levels from different tests. The measurements were collected using the three light fixture positions. The fixture was turned off, 16 sensors were moved to the final set of locations and the measurements were repeated. The entire set of measurements was repeated three times with the sensors rotated among the locations, so that measurements were collected three times with different sensors at each location for each UVGI fixture position (153 data points).
To prevent exposure of personnel to UV-C, the exterior of the patient compartment windows were covered with aluminum foil, warning signage was placed on the ambulance doors and access to the ambulance was restricted during experiments. One person was allowed to enter the ambulance to move the UVGI fixture while it was on; this individual was completely covered in protective clothing and wore a UV-C absorbing face shield.
To examine the effects of surface reflectivity, three different ambulance interior surface conditions were tested. The first condition (called “Original”) consisted of the ambulance interior surfaces as originally supplied by the ambulance manufacturer. The side panels and ceiling were primarily white melamine with aluminum trim and vinyl-covered padding on the edges. The cabinet sliding doors were clear plastic, the access doors were largely covered in diamond-plate aluminum, and the floor was a dark grey non-skid plastic material. Seat cushions were covered with blue vinyl.
For the second interior surface condition (called “Reflective”), many of the vertical white melamine surfaces were covered in diamond-plate aluminum sheets to increase the surface reflectivity for UV light within the patient compartment and thereby increase the irradiation of shadowed areas. Diamond-plate aluminum sheets also were added to the floor beneath the rearfacing primary seat and the patient cot release handle. The positions of the added reflective material are shown in the
For the interior surface condition (called “UV paint”), the added aluminum sheets on the walls were removed, while the original aluminum surfaces and the aluminum added beneath the rearfacing primary seat and the patient cot release handle were left in place. All of the white melamine interior surfaces, including the walls, ceiling and counters, were then painted with UV-C reflecting paint (UVC-MAX, product # W15984, Lumacept, Inc.). The floor, the original aluminum sheets and trim, and the seats and padding were not painted and remained in their original state.
The antimicrobial capabilities of the UVGI system were tested using
The EPA defines a disinfectant as “a substance, or mixture of substances that destroys or irreversibly inactivates bacteria, fungi and viruses, but not necessarily bacterial spores, in the inanimate environment.(
To determine the dose-response relationship between UV-C and
After exposure, each coupon was placed in a 6-well culture plate. Five ml of sterile 0.1% Tween 20 solution (Fisher Scientific) was placed in each well and the plates were shaken for 15 minutes using a titer-plate shaker (Model 4625, Thermo Scientific) at intensity level 5. Twenty μl of each control sample and 100 μl of each UV-exposed sample were then plated onto tryptic soy agar plates (236950, Becton Dickinson) in an exponential spiral using an automated plater (Autoplate, Spiral Biotech). The plates were incubated overnight at 30 °C and colony-forming units were enumerated using an automated plate counter (Qcount Model 510, Spiral Biotech).
The UVGI irradiance (intensity) and spore inactivation data were fitted to an exponential decay model of the form (
S = fraction of spores surviving after UV-C exposure
k = rate constant (cm2/mJ)
I = UV-C irradiance (mW/cm2)
t = exposure time (seconds)
The UVGI dose in mJ/cm2 was determined by integrating the irradiance over the exposure time. The UVGI dose required to disinfect a surface was calculated based upon the k-value found in the laboratory exposure experiments. This surface disinfection dose was divided by the irradiance measurements in the ambulance to calculate the exposure time required to disinfect each location in the compartment (the disinfection time) with the different fixture positions and surface reflectivities. The disinfection dose was used to convert irradiance levels to disinfection times when presenting the results because the disinfection time is easier for the reader to interpret. Irradiance levels for all experiments are provided in the
To test the ability of the UVGI system to inactivate microorganisms in the ambulance, spore-loaded coupons were fastened next to the inlet domes of the ultraviolet light sensors. Ten coupons were placed in the locations in the ambulance with the lowest irradiance levels for each interior surface type. Two control coupons were not exposed to UVGI but were otherwise treated in the same manner. The test was performed 6 times with the original interior surfaces, 3 times with the reflective interior surfaces, and 3 times with the UV-reflective paint on the interior surfaces. For all of these experiments, the UVGI fixture was placed in the back position for first half of the exposure time and in the front position for the second half.
The dose-response curve for
The shortest, median and longest exposure times required for disinfection among the different locations in the ambulance are shown in
The disinfection times when the UVGI fixture was in different positions in the ambulance and with the original interior surfaces are shown in
The effects of adding aluminum sheets or UV-reflective paint to the interior surfaces are shown in
The survival fraction vs. UVGI dose for
Ultraviolet germicidal irradiation is being marketed as a terminal disinfection method for ambulance patient compartments. Commercial portable UVGI systems range in price from around $600 to $3800, depending upon the size and power of the fixture and features such as timers, dose measurement systems and remote controls. UVGI is very effective at inactivating microorganisms if the microbes receive a large enough dose of ultraviolet light. However, ensuring that all surfaces within the compartment are sufficiently irradiated to disinfect them requires careful planning and testing and an appreciation of the limitations of UVGI systems.
UVGI irradiation levels varied greatly from place to place within the ambulance compartment. As seen in
For each combination of UVGI fixture position and surface reflectivity, the median disinfection time among all locations was much less than the longest time (
The position of the UVGI fixture had a substantial effect on the overall disinfection time for the compartment. With the original interior surfaces, moving the fixture from the front to the middle position reduced the disinfection time from 16.5 hours to 3.6 hours. The ratio of the longest time to the median time was also reduced from 148 to 31, suggesting that the light was more evenly distributed. Similar results were seen with the reflective material and the UV-reflective paint. Thus, determining the optimal location for the UVGI fixture is important when designing a UVGI system for an ambulance.
An additional step that can be taken to better distribute the UV light within the compartment is to move the UVGI fixture during disinfection. We examined two possibilities: placing the fixture in the front position for half of the disinfection time and the back position for the other half; and placing the UVGI fixture in the front, middle and back position for one-third of the disinfection time each. With original interior surfaces, moving the fixture did not improve the disinfection time compared to leaving the fixture in the middle position. However, with the reflective aluminum, the disinfection time was reduced from 156 minutes with the light in the front position to 81 minutes by splitting the exposure time between the front and back positions. With the UV-reflective paint, splitting the exposure time between the front and the back reduced the disinfection time from 79 minutes to 67 minutes compared to leaving the fixture in the middle, and splitting the time between the front, middle and back reduced the time to 59 minutes. Further reductions in the disinfection time could be achieved by using a more powerful UVGI fixture or using multiple fixtures for disinfection; using two fixtures, for example, would potentially cut the overall disinfection time in half.
One way to enhance the efficacy of UVGI fixtures is to increase the UV reflectivity of surfaces in the compartment so that more UV irradiation reaches locations that are not directly exposed. Most common materials reflect only a small amount of the UV-C. For example, ordinary white water-based paint is reported to reflect about 10--23% of UV-C light, while stainless steel reflects about 28% and glass only reflects about 4%.(
The UV-reflective paint reduced the disinfection times for all individual locations with all fixture positions, and gave the lowest overall disinfection times. The use of reflective surfaces was not quite as effective, probably because more surfaces were covered with the UV-reflective paint than with the aluminum sheeting, most notably the ceiling. The UV-reflective paint also provides a diffuse reflection, which may help distribute the UV light better, while the reflection from the aluminum sheets is specular (mirror-like). More surfaces could have been covered with the aluminum sheets, but the aluminum sheets have the disadvantage of creating visible light reflections and glare in the compartment. This might be reduced by using aluminum with a brushed or matte finish, but probably can’t be eliminated entirely. On the other hand, the UV-reflective paint that we used is intended for hospital rooms and will not tolerate the repeated scrubbings that need to be performed on some surfaces in the ambulance compartment. Thus, a combination of aluminum sheeting and UV-reflective paint might provide the best method to optimize the performance of a UVGI system.
Our study has several limitations. First, and most important, it is not feasible to perform UV irradiance measurements at every possible location in the ambulance. We attempted to find and measure the locations that received the lowest amounts of UV irradiation, but it is entirely possible that some unmeasured places received less. It is also important to note that covered locations, such as underneath seat cushions or behind cabinet doors, receive no UV exposure at all.
Second, the disinfection time in our paper is based on the use of
Third, the aluminum sheets and UV-reflective paint were freshly applied before our experiments. In a real-world application, these materials would likely lose some reflectivity over time due to dirt, corrosion, abrasion, and other factors.
Fourth, we did not examine the effects of soiling such as by blood droplets or fecal matter on the efficacy of the UVGI system. Such opaque materials are known to block UV-C, and thus it is important that the ambulance be cleaned of any visible contamination before using UVGI as a final step in the disinfection process. Similarly, materials such as patient cots, sheets, papers, clipboards, clothing and other items should be removed from the patient compartment before beginning a UV disinfection cycle, since they will also shield surfaces from UV light.
Finally, although our general conclusions should be widely applicable, the quantitative results in our study are specific to the ambulance compartment that we tested. The irradiance levels, disinfection times, worst locations and other results that we found could be substantially different in ambulances with different UVGI fixtures, cabinet and seat arrangements, surface materials, etc. Thus, any ambulance UVGI system needs to be validated with the actual ambulance configuration for which it is to be used.
Ultraviolet germicidal irradiation systems are potentially a useful tool for reducing the likelihood of infectious disease transmission in ambulance patient compartments. However, when implementing a UVGI terminal disinfection system in ambulances, the following points must be kept in mind:
The amount of UV irradiation delivered to different surface locations can be expected to vary tremendously. When evaluating a UVGI system, multiple locations should be tested.
The time required to disinfect an ambulance compartment is governed by the exposure time needed for the least-irradiated surfaces.
Covered and concealed locations, such as underneath seat cushions or behind cabinet doors, will not be disinfected by a UVGI system.
The position of the UVGI fixture can have a substantial effect on the disinfection time.
Moving the UVGI fixture to multiple locations during a disinfection cycle or using multiple fixtures can reduce the disinfection time.
Increasing the UV reflectivity of interior surfaces can also reduce the disinfection time.
Before putting it into service, any ambulance UVGI system should be thoroughly tested with the actual ambulance configuration for which it is to be used, and should be periodically re-tested to verify that the UVGI system’s performance has not changed over time.
This research was funded by the Office of Public Health Preparedness and Response (OPHPR) of the Centers for Disease Control and Prevention (CDC). The National Institute for Occupational Safety and Health (NIOSH) is a part of the CDC. The findings and conclusions in this report are those of the authors and do not necessarily represent the views of NIOSH. Mention of any company or product does not constitute endorsement by NIOSH.
Ambulance patient compartment showing the locations of the UVGI sensors. A more detailed color schematic and photographs of the ambulance patient compartment are provided in the
Surface disinfection time (exposure time required to inactivate 99.9% of
Effect of UVGI light position on the surface disinfection time at different locations. These experiments were conducted with the original interior surfaces and the UVGI fixture in the front, middle, or back position. The “Frt/Bck” data was calculated assuming that the fixture was in the front position for half of the exposure time and was moved to the back position for the other half. The “Fr/Md/Bk” data was calculated assuming that the fixture was in the front, middle and back positions for 1/3 of the exposure time each. The locations shown were chosen based on the 5 longest disinfection times for each light position. Each bar shows the average of 3 experiments. Error bars show the standard deviation.
Effect of ambulance interior surface treatments on the surface disinfection time. These results were calculated assuming that the UVGI fixture was placed in the front, middle and back positions for one-third of the exposure time each. As described in the materials and methods section, original refers to the ambulance interior surfaces as they were originally made by the manufacturer. Reflective refers to the addition of reflective aluminum plating to various interior surfaces as shown in the
Survival of
Descriptive list of UVGI sensor locations in the ambulance patient compartment. The locations are shown schematically in
| Sensor ID | Sensor location |
|---|---|
| B1 | Under center of high white cabinet on right side of ambulance, facing down |
| B2 | Above center of back cushion on squad bench seat, facing out |
| B4 | On light switch under cabinet by rear-facing primary patient care seat, facing down |
| B5 | On right wall of passageway to cab, facing left, straight across from R1 |
| B6 | Under right side of rear-facing primary care seat between seat and wall, facing wall |
| B7 | Under front edge of rear-facing primary patient care seat, facing floor. |
| B8 | On floor centered in front of secondary patient care seat, facing ceiling |
| B10 | Vertical hand rail on right rear door, facing right wall |
| B11 | Horizontal hand rail on left rear door, facing door |
| B12 | Vertical hand rail on left rear door, facing door |
| B13 | Front of overhead hand rail, facing ceiling |
| B14 | Horizontal hand rail of side door, facing door |
| B15 | Underneath left side of rear-facing primary patient care seat, facing side door |
| G1 | On forward wall next to side door, facing back |
| G2 | On biohazardous sharps disposal box, facing forward |
| G3 | High on right side of passageway to cab, facing left |
| G4 | Next to middle seat belt buckle on front of right-wide squad bench, facing left |
| G5 | Next to oxygen connections on right wall beside side door, facing left |
| G6 | Left wall beside cabinet doors, facing right |
| G7 | Behind horizontal cushion on secondary patient care seat, facing up |
| G8 | On left side of secondary patient care seat alcove, facing back |
| G9 | On side door step, facing door |
| G10 | On ceiling above rear-facing primary patient care seat, facing floor |
| R1 | Next to door of trash bin in passageway to cab, facing right |
| R2 | On upper right cushion next to secondary patient care seat, facing forward |
| R3 | On top of control panel next to rear-facing primary patient care seat, facing right |
| R4 | On work counter in front of control panel next to rear-facing primary care seat, facing up |
| R5 | On latch of storage bin above rear-facing primary patient care seat, facing back |
| R6 | On latch of rearmost cabinet on left side, facing right |
| R7 | In front of cushion for squad bench, facing up |
| R8 | Next to rear-most upper cabinet on right side, facing left |
| R9 | On floor in front of right-side squad bench seat, next to rear door, facing up |
| R10 | On wall above center of rear doors, facing front |
| W1 | On bottom step at side door, back left comer, facing up |
| W2 | On squad bench above biohazardous sharps disposal box, facing forward |
| W3 | Center of back surface of right squad bench cushion, facing right wall |
| W4 | On floor below sharps disposal box, 7.5” from right edge of squad bench, facing up |
| W5 | On top of left vertical cushion above secondary patient care seat, back corner, facing |
| W6 | Center of end of right upper cabinet, facing forward |
| Y1 | On latch of left rear door, facing forward |
| Y2 | On handle of patient cot floor lock, facing floor |
| Y3 | Next to rear-most seat belt buckle on front of right-side squad bench, facing left |
| Y4 | Above control switches on control panel next to rear-facing primary care seat, facing |
| Y5 | On horizontal cushion of rear-facing primary patient care seat, facing up |
| Y6 | On latch of side door, facing left |
| Y7 | Under right corner of rear-facing primary patient care seat, facing ceiling |
| Y8 | Behind cushion for left-side secondary patient care seat, facing up |
| Y9 | Next to latch for roll-up door on front right storage cabinet, facing back |
| Y10 | On back of overhead hand rail, facing ceiling |
Shortest, median and longest disinfection times among the 49 test locations for each interior surface and UVGI fixture position. Note that the shortest time is given in seconds, while the other times are given in minutes. The longest disinfection time is the time required to inactivate 99.9% of the
| UVGI fixture position | Interior surfaces | Shortest disinfection time | Median | Longest disinfection | Location corresponding to longest |
|---|---|---|---|---|---|
| Front | Original | 44 | 6.7 | 990 | B6 |
| Aluminum sheets | 42 | 6.3 | 156 | B6 | |
| UV-reflective paint | 32 | 3.6 | 117 | B6 | |
| Middle | Original | 20 | 7.0 | 216 | B6 |
| Aluminum sheets | 18 | 5.7 | 179 | W1 | |
| UV-reflective paint | 15 | 3.9 | 79 | W1 | |
| Back | Original | 16 | 15.6 | 900 | W5 |
| Aluminum sheets | 17 | 12.1 | 963 | W5 | |
| UV-reflective paint | 14 | 6.8 | 181 | G9 | |
| 1/2 Front | Original | 29 | 5.0 | 244 | B6 |
| Aluminum sheets | 32 | 4.8 | 81 | B13 | |
| UV-reflective paint | 26 | 3.2 | 67 | B6 | |
| 1/3 Front | Original | 37 | 6.9 | 234 | B6 |
| Aluminum sheets | 39 | 5.0 | 79 | W5 | |
| UV-reflective paint | 32 | 3.0 | 59 | B6 |