Conceived and designed the experiments: EMF RES. Performed the experiments: JLW EMF. Analyzed the data: EMF JLW RES. Wrote the paper: EMF JLW RES.
Reusing filtering facepiece respirators (FFRs) has been suggested as a strategy to conserve available supplies for home and healthcare environments during an influenza pandemic. For reuse to be possible, used FFRs must be decontaminated before redonning to reduce the risk of virus transmission; however, there are no approved methods for FFR decontamination. An effective method must reduce the microbial threat, maintain the function of the FFR, and present no residual chemical hazard. The method should be readily available, inexpensive and easily implemented by healthcare workers and the general public. Many of the general decontamination protocols used in healthcare and home settings are unable to address all of the desired qualities of an efficient FFR decontamination protocol. The goal of this study is to evaluate the use of two commercially available steam bags, marketed to the public for disinfecting infant feeding equipment, for FFR decontamination. The FFRs were decontaminated with microwave generated steam following the manufacturers' instructions then evaluated for water absorption and filtration efficiency for up to three steam exposures. Water absorption of the FFR was found to be model specific as FFRs constructed with hydrophilic materials absorbed more water. The steam had little effect on FFR performance as filtration efficiency of the treated FFRs remained above 95%. The decontamination efficacy of the steam bag was assessed using bacteriophage MS2 as a surrogate for a pathogenic virus. The tested steam bags were found to be 99.9% effective for inactivating MS2 on FFRs; however, more research is required to determine the effectiveness against respiratory pathogens.
The potential reuse of National Institute for Occupational Safety and Health (NIOSH) -certified N95 filtering facepiece respirators (FFRs) has been suggested as a possible strategy to conserve available supplies for home and healthcare environments during an influenza pandemic
Some of the previously examined methods, although promising in laboratory studies, may not be universally suited for both healthcare and home environments. Methods that require decontamination equipment such as UV lights, vaporous hydrogen peroxide generators, and moist heat incubators would be better suited for healthcare facilities where such disinfection equipment is more likely to be available. Home environments lack sophisticated decontamination technology, but have disinfectants such as bleach and peroxide; however, the use of these products for FFR decontamination would require customized procedures which may not be easily executed by the general public. Moreover, skin and inhalation health hazards from the use of chemically treated FFRs are a concern
One possibility of overcoming the problems posed by the lack of decontamination equipment and elaborate protocols is to use technology that is readily available and already used by the general public for other applications with similar requirements. Off-the-shelf microwave steam bags (MSBs) are one option that may be used in healthcare and home environments. These bags, typically used to decontaminate breast pump and infant feeding accessories, are available for purchase in many retail stores where infant associated goods are sold. The instructions are written for the general public and are based on the operation of a microwave oven, which are readily available in home and healthcare environments. The use of commercially available steam bags for FFR decontamination has not been investigated, although previous studies suggest that microwave generated steam decontamination is promising
| FFR Details | Water Content (g) | Filtration Efficiency (%) | ||||
| Model | Type | Contains Hydrophilic layer(s) | After decon. | 60 min | As received | After 1X |
| 3M 1870 | Surgical | no | 0.4 | 0.1 | 99.67 | 99.62 |
| 3M 1860 | Surgical | yes | 13.5 | 9.6 | 99.28 | 99.47 |
| KC PFR95 | Surgical | no | 0.9 | 0 | 96.13 | 95.77 |
| 3M 8210 | Particulate | yes | 11.6 | 8.2 | 99.88 | 99.34 |
| Cardinal Health | Particulate | yes | 12.8 | 11.2 | 99.62 | 99.56 |
| Moldex 2200 | Particulate | no | 1.5 | 0.2 | 98.52 | 99.24 |
*Data modified from references (10) and (14).
Determined using MSB X bags.
n = 1.
In the second phase of testing, the triplicate samples for each of the FFR models, 3M 1870, Kimberly-Clark PFR95, and Moldex 2200, passed the filtration efficiency testing after three cycles of decontamination using both steam bag brands (
| FFR | Water Content (g) | Filtration Efficiency (%) | ||||
| Type | Model | After (3X) decon. | 30 min | As received | MSB X (3X) | MSB Y (3X) |
| Surgical | 3M 1870 | 1.7±1.4 | 0.1±0.1 | 99.7±0.1 | 98.6±0.6 | 99.0±1.1 |
| Surgical | KC PFR95 | 1.3±1.3 | 0±0 | 96.1±0.4 | 95.5±0.3 | 96.4±1.2 |
| Particulate | Moldex 2200 | 0.9±0.4 | 0.1±0.1 | 98.5±1.0 | 98.6±0.8 | 98.4±1.5 |
Determined using MSB X bags.
n = 3.
| FFR Model | MS2 from load controls | CV (%) | MS2 from Steam Treated FFR | Difference (Load vs. Treated) | Reduction (%) | |
| 7.57±0.08 | 18.5 | 4.47±0.32 | 3.10 | 99.90 | ||
| 7.09±0.17 | 37.6 | 3.85±0.35 | 3.25 | 99.93 | ||
| 9.96±0.06 | 14.6 | 5.32±0.30 | 4.64 | 99.99 | ||
| 6.93±0.16 | 32.8 | ≤3.69# | ≥3.24 | ≥99.94 | ||
| 8.15±0.25 | 62.3 | 4.70±0.69 | 3.45 | 99.93 | ||
| 7.04±0.09 | 19.8 | ≤3.93# | ≥3.11 | ≥99.86 |
* Values in Log10 (pfu/FFR).
# Two of three trials reached detection limits.
Commercially available MSBs offer intrinsic benefits for FFR decontamination in home and healthcare settings. The steam bags, constructed for the purpose of disinfection (baby bottles and breast pumps), are readily available for purchase. The instructions for use are clearly provided on the side of the bags (
Top: MSB X bags. Bottom: MSB Y bags.
Steam sterilization, by the use of autoclaves, is routinely used in the processing of medical equipment. Autoclaves produce high pressure saturated steam and are effective at inactivating microorganisms including spores
The FFR filtration performance for the three cycle treatments was within acceptable levels of the selection criterion for each FFR model treated in each steam bag brand. Bergman et al. reported similar results with no deleterious effect of microwave generated steam on the filtration performance of three surgical and three particulate N95 FFRs
The use of a steam bag has some limitations for FFR decontamination. The steam bags do not compartmentalize the water reservoir and sample location. The FFR is placed directly into the water in the reservoir, which produces the potential for water absorption by the FFR material. Water absorbency is important as a saturated FFR would require an extended drying period before reuse is possible. An extended drying period is counterproductive to increasing FFR supply in the event of shortages due to high demand. The potential to use the steam bags for FFR decontamination will likely be FFR model specific, as demonstrated by the water absorption data in this study (
The decontamination procedure and demonstrated efficacy of the MSBs are not in alignment with current FDA guidelines and requirements for the reuse of single use medical devices
A cleaning procedure was not included as part of this study; however, it was previously demonstrated that soil load accumulation may not significantly impact microwave generated steam decontamination of FFRs
In healthcare and home environments alike, the performance of the steam, and microwave ovens, may demonstrate some variability. It should be noted that the microwave used in this study was rated at 1100 W by the manufacturer, but was experimentally determined to function at 750 W previously
More studies are required before the use of steam bags can be considered for FFR decontamination for the purpose of reuse. In general, reuse requires a higher degree of rigor than single use applications. Commonly used in the evaluation of medical devices, an ultrastructural analysis of the decontaminated FFRs may help to address concerns and knowledge gaps associated with FFR reuse. Quality control assessments of the steam bags and microwave ovens should be performed to investigate the utility of the steam bag decontamination procedure. Likewise, implementing MSB decontamination of FFRs in home and healthcare settings would present quality control issues which should be investigated for each environment. The use of MS2, a nonenveloped virus, in this study does not accurately reflect the potential efficacy of the steam bag against enveloped viruses including 2009 H1N1. In general, enveloped viruses are more susceptible to decontamination due to the fragile lipid coat. Testing the steam bags against other microbes can assist discerning the steam bags' potential. FFRs decontaminated using the steam bags should be fit tested to ascertain if any changes to FFR shape and fit occurred as a result of the steam process, although previous research with microwave generated steam suggests that this is unlikely.
The evaluation of the feasibility of steam bag decontamination of FFRs was studied in three phases. In the first phase, a preliminary screening of the six models of respirators treated in one brand of MSB was conducted using two quality objectives: limited filtration performance degradation and low water absorbency/retention. Each quality objective was evaluated using a predetermined standard. Firstly, the steam bag decontamination must not degrade the filtration performance of the FFR below the efficiency required (95% efficient) by NIOSH certification requirements outlined in 42 CFR 84. Secondly, the FFR must be dry (defined for this study as less than 1 g water content) within 60 min of drying time under room conditions (approx. 20°C and 60% RH). The rationale for this requirement is the users would be unlikely to find wearing a wet respirator to be uncomfortable, previously identified as a barrier to respirator tolerability
In phase 2 testing, FFR models passing the preliminary assessment were evaluated for filtration efficiency following three cycles of steam bag sterilization which included a 30 min drying period between treatments. Each model was evaluated in triplicate for each MSB brand. The FFRs were evaluated for water absorption/retention after 30 min of drying time following steam treatment using one MSB brand. FFR models exceeding the predefined quality standards were eliminated from phase 3 testing.
In the final phase of testing, the decontamination efficacy of the steam bag was determined in triplicate for the FFR models passing the phase 2 evaluation using both brands of bags. For each FFR model, six samples were contaminated with MS2 droplets. The MS2 from three of the samples for each FFR model was collected and enumerated to determine the loading level. The other three samples were decontaminated using one brand of MSBs. The process for each FFR was repeated for the second brand of MSBs. Upon decontamination the MS2 was collected from the filter samples and enumerated via plaque assay.
Six respirator models were used in this study. Three of the models, namely the 3M 1870 (3M, St. Paul MN), 3M 1860 (3M, St. Paul MN), and the Kimberly-Clark PFR95 (Kimberly-Clark, Dallas, TX) are surgical N95 FFRs. Surgical N95 FFRs are NIOSH-approved particulate respirators that have also been cleared by the Food and Drug Administration (FDA) as medical devices. Three particulate FFR models included in the study are the 3M 8210 (3M, St. Paul MN), Moldex 2200 (Moldex, Culver City, CA) and Cardinal Health (Cardinal Health, Dublin, OH). All models used were available in the Strategic National Stockpile at the time of writing except for the Cardinal Health FFR, which was randomly selected from the laboratory stock.
The manufacturer's instructions for use with baby feeding accessories were applied to the steam treatment of FFRs. The instructions were the same for each steam bag brand. Individual FFRs were placed inside separate bags filled with 60 ml of tap water (
A Model 8130 Automated Filter Tester (AFT) (TSI, Inc., St. Paul, MN, USA) was used to measure initial percent filter aerosol penetration and filter airflow resistance for FFR models as received (control), 1 cycle treated FFRs, and 3 cycle treated FFRs. The TSI 8130 AFT delivers a solid polydispersed sodium chloride (NaCl) aerosol that meets the particle size distribution criteria set forth in 42 CFR 84 Subpart K, Section 84.181 for NIOSH certification (CFR, 1995). Filter penetration testing was performed using a similar but abbreviated version of the NIOSH certification protocol previously used to evaluate FFR filtration performance
FFRs were decontaminated using the MSB X bags as described above. The FFRs were weighed prior to decontamination to determine the dry weight and reweighed immediately following decontamination and after a predetermined drying period of 30 or 60 min to determine the wet weight. The dry weight was subtracted from the wet weight to determine the amount of water absorbed or retained by the material of the FFR. MSB Y bags were not used to assess FFR water absorbency; however, comparable water absorbency values for FFRs treated in the in both brands of bags are expected due to the similar steam bag designs, which have the FFR partially submerged in the water (
The media, virus, and host cells used in this research have been described previously
Virus containing droplets were applied to FFRs using a spray bottle (Fisherbrand Adjustable-Spray Mini-Wash Bottle, Fisher Scientific, Pittsburgh, PA). FFRs were mounted (friction fitted) to a funnel, which served as the FFR holder, and attached to a ring stand (
Filtration efficiencies for the FFR models were calculated from the percent aerosol penetration values (
The repeatability of the MS2 application technique (spray bottle) used to contaminate the FFRs was assessed using ASTM Standard E2721-10
The antiviral activity of the steam bags was determined for each FFR model by comparing the average log10 pfu of MS2 loaded onto three untreated (control) FFR samples with three steam-treated FFR samples. Percent reduction was also calculated for a given FFR model by dividing the number of recovered pfu from a treated respirator by the average pfu recovered from the untreated controls. The quotients calculated using the three treated samples were averaged to give average percent reduction.
The authors wish to express our sincere gratitude to Evanly Vo, Dennis Viscusi, Mike Bergman, and the manuscript reviewers (Lisa Delaney, Brian Heimbuch, and Ziqing Zhuang) for their suggestions and contributions. The findings and conclusions in this manuscript have not been formally disseminated by the NIOSH and should not be construed to represent any agency determination or policy. Mention of any company, product, or policy, or the inclusion of any reference does not constitute endorsement by NIOSH.