10118945831897J Occup Environ HygJ Occup Environ HygJournal of occupational and environmental hygiene1545-96241545-963223339418463611510.1080/15459624.2012.750554HHSPA730975ArticleClosed-Face Filter Cassette (CFC) Sampling—Guidance on Procedures for Inclusion of Material Adhering to Internal Sampler SurfacesAshleyKevin1HarperMartin2U.S. Department of Health and Human Services, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, Cincinnati, OhioU.S. Department of Health and Human Services, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, Morgantown, West VirginiaCorrespondence to: Kevin Ashley, U.S. Department of Health and Human Services, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, 4676 Columbia Parkway, Cincinnati, Ohio 45226; kashley@cdc.gov21102015201306112015103D29D33

The sampling and analytical methods published in the NIOSH Manual of Analytical Methods (NMAM)(1) represent state-of-the-art methods for assessing worker exposures to toxic chemicals. Aerosol sampling methods generally specify the collection of workplace air samples using samplers containing filters such as 37-mm closed-face cassettes (CFCs). NIOSH considers that all particles entering CFCs, as well as some other samplers, such as the Institute of Occupational Medicine (IOM) sampler, should be included as part of the sample whether they deposit on the filter or on the inside surfaces of the sampler. This matter has been discussed in detail in the section on “Sampler Wall Losses” of NMAM Chapter O(2) and has also been specifically addressed in a gravimetric method promulgated by the Occupational Safety and Health Administration (OSHA).(3) The most accurate measure of occupational exposures therefore includes all aerosol particles entering workplace air samplers, both for gravimetric analysis and for analytes such as metals and metalloids. Hence, following sampling, i.e., during sample preparation and analysis, procedures should be used to account for material adhering to the internal surfaces of air sampling cassettes.

Research studies have shown that material deposited on the inside surfaces of air sampling cassettes often constitutes a significant fraction of the aerosol that enters the sampler.(47) Example data for CFCs are presented in Table I,(8) which demonstrate that, in some instances, the wall deposit can even exceed the amount of material collected on the filter. There are several mechanisms that can lead to particles being deposited on internal surfaces rather than on the filter, including direct impaction, gravitational settling, interception during transport by eddies within the sampler, bounce from the filter, and electrostatic attraction. Losses of collected particulate matter from the filter can also occur during sample shipment and handling. It is likely that multiple mechanisms are in play simultaneously, with varying degrees of importance depending on factors such as particle density, inertial velocity, wind speed and orientation, and so on. However, it is unlikely that these mechanisms would interfere with particles entering human airways during breathing.

Recent laboratory and field investigations on airborne particles of up to 20 μm aerodynamic equivalent diameter containing lead (Pb) have shown that there is no qualitative difference in the size distribution of aerosols deposited on the filter or on the internal surfaces of CFCs.(9,10) Both components therefore would contribute to the inhaled dose that is delivered to the airways of exposed workers. This is a concern not only for sampling and analysis of lead and other elements but can apply to aerosol sampling in general, and to certain targeted aerosol fractions (i.e., inhalable, respirable, or thoracic).(11,12) Sampling cassettes composed of conductive materials do minimize the internal deposits that occur through static attraction but will not eliminate internal deposits entirely.(2) Losses of collected particulate matter from the filter during sample transport and handling are also possible, regardless of the composition of the sampler.

Consideration of deposits adhering to internal sampler walls is included in related international voluntary consensus standards published by ASTM International (formerly, American Society for Testing and Materials) and the International Organization for Standardization (ISO), which describe the sampling and analysis of airborne metals and metalloids in occupational atmospheres.(1317) The recommendation to account for sampler wall deposits is consistent with the position of OSHA, whose sampling and analytical methods for metals and metalloids specify transferring loose dust, washing internal surfaces, and wiping internal cassette walls.(18,19)

Depending on the analyte(s), suitable techniques for accounting for internal sampler wall deposits include: (1) wiping (with inclusion of the wipe along with the filter during subsequent sample preparation);(18,19) (2) carrying out sample extraction directly within the cassette;(2022) (3) thorough washing of internal cassette surfaces;(21) and (4) the use of internal sampling capsules or cartridges (sampler inserts).(3,23) These techniques are consistent with NIOSH methodologies that, as noted, should include not just the filter catch but all material that enters the sampler. In some cases an exception might be made where sampling cassettes composed of conductive materials are employed and internal wall deposits are thereby minimized.(2)

NIOSH methods that are presently affected by the recommendation to account for cassette wall deposits are listed in Table II. Specific guidance on techniques to include internal wall deposits is provided below.

Three NIOSH CFC methods entailing gravimetric analysis are listed in Table IIa. When CFC sampling is carried out, gravimetric analysis procedures for aerosol measurement should employ internal capsules(24) such as the polyvinyl chloride (PVC) Accu-Cap (SKC Inc., Eighty Four, Pa.), the Woodcheck (MSA, Pittsburgh, Pa.), or equivalent. Weighing only the filter is generally insufficient, since this practice will not account for material collected on internal sampler walls and on the interior of the cap. (While, in principle, each CFC sampler could be weighed in its entirety, contamination of outside cassette surfaces would contribute greatly to sampling bias, especially in dusty work environments. There is also a loss in precision caused by the much greater absolute mass.) For respirable sampling and gravimetric analysis where cyclone/cassette assemblies are used, internal non-filter deposits are still noted,(7) but it is more difficult to account for them; thus, they should be minimized where possible through the use of cassettes composed of conductive materials, with collection of the aerosol deposits onto pre-weighed filters.(2)

NIOSH CFC sampling and analytical methods for which inclusion of internal non-filter deposits is already described in each method, at least to some extent, are listed in Table IIb. These methods, which entail rinsing, prescribe rinsing if there is visible deposit on internal cassette surfaces (and it is expected that the target analyte is soluble in the solution used for rinsing). The rinsate is then included along with the filter in subsequent sample extraction steps prior to analytical measurement. In fact, rinsing should take place even if there is no visible deposit on the internal walls or inside the cap. Although often invisible to the naked eye, internal non-filter deposits of aerosol particles can contribute significantly to the total amount of aerosol entering the sampler. Another technique that merits consideration is within-cassette extraction, which could prove to be effective for some or all of the methods listed in Table IIb.

Table IIc lists elemental analysis methods, in which there are options for inclusion of CFC internal non-filter deposits. The use of a digestible capsule that is fused to the filter (e.g., an internal cartridge composed of cellulose acetate attached to a mixed-cellulose ester (MCE) filter) is especially appealing as this obviates any need for a separate wiping or rinsing step. After sampling, the capsule can be simply removed from the cassette and placed into a sample preparation vessel for dissolution in its entirety. (Since many digestion vessels do not easily accommodate a 37-mm diameter internal capsule, the fabrication of a 25-mm diameter version is a worthwhile goal.) Cassette wiping, where particles are dislodged using manual pressure through a wetted fabric or similar material, is a practice that OSHA uses for elemental analysis, and the effectiveness of this technique has been demonstrated.(18) The wipe is added to the filter used for sampling, and both are digested together prior to subsequent instrumental analysis. Alternative techniques such as cassette rinsing with dilute acid solution could be considered, whereby the rinsate from the cassette walls is added to the sample preparation vessel containing the filter used for sampling. These practices may be acceptable, but their effectiveness must be validated through demonstrated quantitative recoveries. In cases where extraction and not digestion is called for, it is possible to carry out the extraction directly within the cassette, which automatically incorporates non-filter deposits into the dissolution medium.(21,22,25)

Other methods for which internal non-filter deposits should be included but do not fit into the above three categories (i.e., for methods listed in Table IIa–c) are listed in Table IId. For hexavalent chromium (Cr[VI]) methods, cassette wiping with a non-reactive medium such as PVC can be used. (With Cr[VI]-containing samples, cellulosic wiping materials should not be used since these media can cause reduction of Cr[VI].) This wipe, which may be a filter wetted with extraction solution, is then added to the filter used for sampling prior to further analytical processing. As an alternative, Cr[VI] can be extracted directly within the cassette using the extraction solution (basic buffer),(21) and aliquots can then be analyzed. Still another option could entail rinsing of Cr[VI] wall deposits with basic buffer solution. Rinsing with dilute acid solution should be used for removal of alkaline dusts from internal cassette surfaces; alternatively, a wiping procedure may be employed.

In the future, additional NIOSH methods may require incorporation of internal sampler wall deposits; a number of these methods are listed in Table III. Effective procedures and performance data are needed for all of the listed methods; thus, for the time being, the general recommendation to account for internal sampler wall deposits does not apply to these methods.

For most of the aerosol sampling methods discussed here, alternatives to the CFC samplers are available, and some of these have been optimized or found by experiment to collect the ISO-defined “inhalable fraction.” Several such samplers are available commercially.(2,26) Deposits also occur on internal surfaces of these samplers (e.g., within the cones of the IOM and Gestamtstaubprobenahme (GSP) samplers, and on the Oring of the Button sampler), and the user should be aware of and, if necessary, account for them.

In summary, owing to the significance of internal non-filter sampler deposits, it is generally recommended to account for this contribution to worker exposures to airborne chemical agents. This recommendation may apply to methods that collect and analyze both aerosols and vapor where, for example, the sampler consists of a filter cassette and sorbent tube in series; in such cases, inclusion of the particulate material adhering to the filter cassette internal surfaces should be considered. Inclusion of internal non-filter sampler deposits may not pertain to methods that entail the use of impregnated filters (e.g., isocyanates). Other NIOSH aerosol sampling methods not addressed here (Table III) may deserve consideration of internal wall deposits in the future. It is intended to revise the texts of affected NIOSH methods to specifically account for internal wall deposits. However, until revised methods are approved and published, the recommendations given here shall apply.

Mention of company names or products does not constitute endorsement by the U.S. Department of Health and Human Services, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health.

National Institute for Occupational Safety and Health (NIOSH)NIOSH Manual of Analytical Methods (NMAM)AshleyKO’ConnorPFCincinnati, OhioNIOSHBaronPAFactors affecting aerosol samplingNIOSH Manual of Analytical MethodsChapter OCincinnati, OhioNIOSH2003Occupational Safety and Health Administration (OSHA)Method PV2121OSHA Sampling and Analytical MethodsSandy, UtahOSHA2003DemangeMGendreJCHervé-BazinBCartonBPeltierAAerosol evaluation difficulties due to particle deposition on filter holder inner wallsAnn Occup Hyg343994031990DemangeMGörnerPElcabacheJMWrobelRField comparison of 37-mm closed-face filter cassettes and IOM samplersAppl Occup Environ Hyg17200208200211871756PuskarMAHarkinsJMMoomeyJDHeckerLHInternal wall losses of pharmaceutical dusts during closed-face, 37-mm polystyrene cassette samplingAm Ind Hyg Assoc J5228028619911951066DobsonLReichmannLPoppDEvaluation of quartz residue on cassette interiors of AIHA proficiency samplesJ ASTM Int24162005HarperMDemangeMConcerning sampler wall losses in the chemical analysis of airborne metalsJ Occup Environ Hyg4D81D86200717613720LeeTChisholmWPSlavenJEHarperMSize distributions of 0.5 to 20 μm aerodynamic diameter lead-containing particles from aerosol sampler walls and filtersAerosol Sci Technol43104210502009ChisholmWPLeeTSlavenJENelsonJHarperMComparison of filter and wall deposits from samplers used to collect airborne lead-containing dusts at field sitesAerosol Sci Technol464114182012HarperMA review of workplace aerosol sampling procedures and their relevance to the assessment of beryllium exposureJ Environ Monit8598604200616767225BrissonMJArchuletaMThe real issue with wall deposits in closed-face filter cassettes—What’s the sample?J Occup Environ Hyg6783788200919894181ASTM InternationalASTM D6785Standard Test Method for Determination of Lead in Workplace Air Using Flame or Graphite Furnace Atomic Absorption SpectrometryWest Conshohocken, PaASTM International2008ASTM InternationalASTM D7035Standard Test Method for Determination of Metals and Metalloids in Airborne Particulate Matter by Inductively Coupled Plasma Atomic Emission Spectrometry (ICP-AES)West Conshohocken, PaASTM International2010ASTM InternationalASTM D7439Standard Test Method for Determination of Elements in Airborne Particulate Matter by Inductively Coupled Plasma Mass SpectrometryWest Conshohocken, PaASTM International2008International Organization for Standardization (ISO)ISO 15202Workplace air—Determination of metals and metalloids in airborne particulate matter by inductively coupled plasma atomic emission spectrometry3 PartsGenevaISO2011http://www.iso.orgInternational Organization for Standardization (ISO)ISO 30011Workplace Air—Determination of Metals and Metalloids in Airborne Particulate Matter by Inductively Coupled Plasma Mass SpectrometryGenevaISO2011HendricksWStonesFLillquistDOn wiping the interior walls of 37-mm closed-face cassettes—An OSHA perspectiveJ Occup Environ Hyg6732734200919894173Occupational Safety and Health Administration (OSHA)Methods ID-121, ID-125g and ID-215OSHA Sampling and Analytical MethodsSandy, UtahOSHA2008AshleyKAndrewsRNCavazosLDemangeMUltrasonic extraction as a sample preparation technique for elemental analysis by atomic spectrometryJ Anal At Spectrom16114711532001AshleyKApplegateGTMarcyADEvaluation of sequential extraction procedures for soluble and insoluble hexavalent chromium compounds in workplace air samplesJ Environ Monit11318325200919212588Institut National de Recherche et de Sécurité (INRS): Métaux—Métalloïdes (Fiche 003)Métrologie des Polluants—Évaluation de l’Éxposition Professionelle—Methodes de Prélèvement et d’Analyse de l’AirMeasurement of Pollutants—Evaluation of Occupational Exposures—Methods of Air Sampling and AnalysisParisINRS2008In FrenchHarperMAshleyKPreliminary studies on the use of acid-soluble cellulose acetate internal capsules for workplace metals sampling and analysisJ Occup Environ Hyg9D125D129201222702231VolkweinJCMaynardADHarperMChapter 25: Workplace aerosol measurementAerosol Measurement—Principles, Techniques, and Applications3KulkarniPBaronPAWillekeKNew YorkWiley2011AshleyKAgrawalACroninJTonazziJMcCleskeyTMEhlerDSUltra-trace determination of beryllium in occupational hygiene samples by ammonium bifluoride extraction and fluorescence detection using hydroxybenzoquinoline sulfonateAnal Chim Acta584281286200717386616VincentJHAerosol SamplingNew YorkWiley2007

Closed-Face Filter Cassette Internal Wall Deposits from Samples Obtained in Various Workplace Environments

Work Environment/ActivitynChemical Agent(s)Maximum Wall Deposit(s) (%)Median Wall Deposit(s) (%)
Copper smelter17Pb5521
Lead ore mill28Pb3519
Solder manufacture30Pb7429
Battery production16Pb6628
Welding10Cr[VI]555
Electroplating12Cr[VI]1712
Paint spray29Cr[VI]127
Foundry9Zn6253
Zinc plating18Zn9127
Cast iron foundry18Fe4622
Grey iron foundry18Fe7724
Bronze foundry6Cu, Pb, Zn45, 17, 2119, 13, 15
Cuproberyllium foundry4Cu, Be40, 3931, 12

NIOSH Methods Affected by the Recommendation to Account for Cassette Internal Non-Filter Deposits, and Suitable Techniques for Inclusion of All Aerosol Entering the Sampler

NIOSH MethodsTechnique RecommendedComments
(a) Gravimetric Analysis Methods
0500 - Particles (“total”)Use internal capsule for sample collection, which is analytically weighed pre- and post-samplingSuitable examples are capsules composed of PVC attached to PVC filters
5000 - Carbon black
0600 - Particles (respirable)Use filter housed in static-dissipative (“conductive”) cassette, or use internal capsuleConductive or static-dissipative sampler material can minimize internal wall deposits
(b) Internal Deposits Currently Considered in Method Text
5005 – ThiramRinse internal wall deposits with extraction solution (e.g., water, eluent, organic solvent)Rinsate is added to sample preparation vessel (e.g., beaker) along with the filter used for sampling
5011 - Ethylene thiourea
5030 - Cyanuric acid
5032 - Pentamidine isethionate
5700 - Formaldehyde (textile or wood dust)Use IOM inhalable samplerAlternative procedure could entail CFC sampling, followed by rinsing as above
(c) Elemental Analysis Methods
7013 - Aluminum

Remove internal wall deposits with wetted wipe (e.g., filter or wipe material); wipe is then added to sample preparation vessel (e.g., beaker) along with the filter used for sampling

Use digestible internal capsule (e.g., cellulose acetate attached to MCE filter) for sample collection; capsule is digested and analyzed in its entirety

Alternative procedure could entail rinsing of internal sampling cassette surfaces with dilute acid solution (demonstrate effectiveness)
7024 - Chromium
7027 - Cobalt
7029 - Copper
7030 - Zinc
7048 - Cadmium
7046 - Barium
7074 - Tungsten
7082 and 7105 - Lead
7102 - Beryllium
7300, 7301, and 7303 - Elements
7900 - Arsenic
7020 - CalciumRemove internal wall deposits with wetted filter or wipe, which is then added to sample preparation vessel (e.g., centrifuge tube) along with the filter used for samplingAlternative procedures involve within-cassette extraction
7701 - Lead
7704 - Beryllium
(d) Hexavalent Chromium and Alkaline Dusts
7600, 7605, and 7703 - Hexavalent chromiumRemove internal wall deposits with wetted filter, which is then added to sample preparation vessel (e.g., beaker) along with the filter used for samplingAlternative procedures could involve rinsing or within-cassette extraction
7401 - Alkaline dusts

Other NIOSH Methods Potentially Affected by the Recommendation to Account for Internal Sampler Wall Deposits

NIOSH Methods and Analytes
5001 - 2,4-D and 2,4,5-T ((2,4-dichlorophenoxy)acetic acid and 2,4,5-trichlorophenoxy)acetic acid)
5002 - Warfarin
5003 - Paraquat
5004 - Hydroquinone
5006 - Carbaryl
5007 - Rotenone
5008 - Pyrethrum
5009 - Benzoyl peroxide
5010 - Bromoxynil and Bromoxynil octanoate
5012 - EPN (Ethyl p-nitrophenyl thionobenzenephosphonate)
5013 - Benzidine, o-Tolidine, and o-Dianisidine dyes
5014 - Chlorinated terphenyl
5016 - Strychnine
5017 - Dibutyl phosphate
5018 - 2,4,7-Trinitrofluoren-9-one
5019 - Azaleic acid
5020 - Dibutyl phthalate and Di(2-ethylhexyl) phthalate
5021 - o-Terphenyl
5022 - Organo-arsenic
5023 - Coal tar pitch volatiles
5025 - Chlorinated diphenyl oxide
5026 - Oil mist
5027 - Ribavirin
5031 - Aspartame
5033 - p-Nitroaniline
5034 - Tributyl phosphate
5035 - Super-absorbent polymers
5036 - Trimetallic anhydride
5037 - Triorthocresyl phosphate
5038 - Triphenyl phosphate
5039 - Chlorinated camphene
5041 - Capsaicin and Dihydrocapsaicin
5042 - Asphalt fume (benzene soluble and total particulate)
5043 - p-Toluenesulfonic acid
5044 - Estrogenic compounds
5516 - 2,4 and 2,6 Toluenediamine
5521, 5522, 5525 - Isocyanates
5524 - Metalworking fluids
5527 - Triphenyltin chloride
7500, 7501, 7601, 7602, and 7603 - Silica