87018586404Pediatr Infect Dis JPediatr. Infect. Dis. J.The Pediatric infectious disease journal0891-36681532-098724346605460692710.1097/INF.0000000000000089HHSPA726259ArticleChallenges With New Rapid Influenza Diagnostic TestsOlsenSonja J.PhDInfluenza Division, Centers for Disease Control and Prevention Atlanta, GAInfluenza Program, Thailand MOPH – US CDC Collaboration Nonthaburi, ThailandKittikraisakWanitchayaPhDInfluenza Program, Thailand MOPH – US CDC Collaboration, Nonthaburi, ThailandFernandezStefanPhDArmed Forces Research Institute of Medical Sciences, Bangkok, ThailandSuntarattiwongPiyaratMDQueen Sirikit National Institute of Child Health, Bangkok, ThailandChotpitayasunondhTaweeMDQueen Sirikit National Institute of Child Health, Bangkok, Thailand29920151201415102015331117118To the Editors

Rapid influenza diagnostic tests (RIDTs) are often used at point-of-care due to their ease of use and rapidly available results. Most tests are lateral flow immunoassays that detect chromatographic changes if an influenza antigen is present in the respiratory specimen. These tests have high specificity (therefore, a positive is almost certainly a true positive) but low sensitivity (therefore, will often miss true cases).1,2 A newer immunofluorescence assay, Sofia A+B FIA (Quidel, San Diego, CA), demonstrated increased sensitivity but maintained high specificity.3 However, on December 3, 2012, Quidel issued a voluntary recall of certain lots of Sofia A+B because of false positive results.4

In August 2011, we began a prospective cohort study of children aged ≤36 months at Queen Sirikit National Institute of Child Health, the largest pediatric referral hospital in Thailand. Children (equal numbers of high risk and healthy) are followed for 2 years and parents contacted weekly to inquire about whether their child had acute respiratory illness. Ill children came to the hospital and had a combined nasal and throat swab collected and tested for influenza viruses by realtime reverse transcription polymerase chain reaction (rRT-PCR).5 In addition, a separate nasal swab was taken and tested using 1 of 2 RIDTs made by Quidel (QuickVue A+B during August 2011 to January 20, 2013; Sofia A+B during January 21, 2013 to May 2013).

Of the 1152 specimens tested with QuickVue A+B, 59 (5.1%) were positive by rRT-PCR. Compared with rRT-PCR, Quick-Vue A+B had a sensitivity of 55.9% (33/59; 95% confidence interval (CI): 42.4–68.8%) and a specificity of 99.4% (1086/1093; 95% CI: 98.7–99.7%). Seven (0.6%) were false positive on QuickVue A+B. Of the 370 specimens tested with Sofia A+B, 12 (3.2%) were positive by rRT-PCR. Compared with rRT-PCR, Sofia had a sensitivity of 100% (12/12; 95% CI: 73.5–100.0%) and a specificity of 61.2% (219/358; 95% CI: 55.9–66.3%). One-hundred thirty nine (38.8%) were false positive on Sofia. Of the 139 false positives, 123 (88.5%) were influenza B and 16 were influenza A. There was no difference in the time between illness onset and specimen collection date between true and false positives (median days = 2; P = 0.96), nor was there a difference between the prevalence of influenza during the 2 periods (5.1% vs. 3.2%, P = 0.1). Two lots of the Sofia assay were used and both had poor specificity (data not shown).

We learned that the assay lots we purchased were made after the recognized problem that led to the recall was corrected. Nevertheless, our data suggest that continuing problems exist. We share these data to alert others using this assay to the risk of false positives.

Our findings further support the limitations in using RIDTs for clinical decision making.6 In Thailand, the Clinical Practice Guidelines recommend oseltamivir treatment for 2 groups of patients: (1) patients with complicated or severe influenza (ie, pneumonia, alteration of conscious, loss of appetite or dehydration and symptoms worsening after 48 hours of illness) and (2) patients with a high risk of having severe influenza (ie, pregnancy, obesity or chronic medical conditions). They also state that oseltamivir should also be considered in children <2 years of age or persons >65 years of age. In Thailand, the use of RIDTs is not recommended for deciding on a course of treatment. Further, we caution that newer RIDTs using immunofluorescence may need additional field evaluations.

The findings and conclusions in this report are those of the authors and do not necessarily represent the views of the Centers for Disease Control and Prevention. The views and opinions expressed herein are the private opinions of the authors and do not reflect necessarily those of the US Army or the Department of Defense. The authors have no funding or conlflicts of interest to disclose.

ChartrandCLeeflangMMMinionJAccuracy of rapid influenza diagnostic tests: a meta-analysisAnn Intern Med201215650051122371850Centers for Disease Control and PreventionEvaluation of 11 commercially available rapid influenza diagnostic tests–United States, 2011–2012MMWR Morb Mortal Wkly Rep2012614387387623114254LewandrowskiKTameriusJMenegusMDetection of influenza A and B viruses with the Sofia analyzer: a novel, rapid immunofluorescence-based in vitro diagnostic deviceAm J Clin Pathol201313968468923596120US Federal Drug AdministrationClass 2 recall Sofia influenza A + B FIA, Kit #202182013Available at: http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?ID=115438. Accessed July 15, 2013Centers for Disease Control and PreventionCDC protocol of realtime RTPCR for influenza A (H1N1)2009Available at: http://www.who.int/csr/resources/publications/swineflu/realtimept-pcr/en/index.html. Accessed March 3, 2010Centers for Disease Control and PreventionGuidance for clinicians on the use of rapid influenza diagnostic tests2013Available at: http://www.cdc.gov/flu/professionals/diagnosis/clinician_guidance_ridt.htm. Accessed July 15, 2013