Rapid influenza diagnostic tests (RIDTs) are often used at point-of-care due to their ease of use and rapidly available results. Most tests are lateral flow immunoassays that detect chromatographic changes if an influenza antigen is present in the respiratory specimen. These tests have high specificity (therefore, a positive is almost certainly a true positive) but low sensitivity (therefore, will often miss true cases).
In August 2011, we began a prospective cohort study of children aged ≤36 months at Queen Sirikit National Institute of Child Health, the largest pediatric referral hospital in Thailand. Children (equal numbers of high risk and healthy) are followed for 2 years and parents contacted weekly to inquire about whether their child had acute respiratory illness. Ill children came to the hospital and had a combined nasal and throat swab collected and tested for influenza viruses by realtime reverse transcription polymerase chain reaction (rRT-PCR).
Of the 1152 specimens tested with QuickVue A+B, 59 (5.1%) were positive by rRT-PCR. Compared with rRT-PCR, Quick-Vue A+B had a sensitivity of 55.9% (33/59; 95% confidence interval (CI): 42.4–68.8%) and a specificity of 99.4% (1086/1093; 95% CI: 98.7–99.7%). Seven (0.6%) were false positive on QuickVue A+B. Of the 370 specimens tested with Sofia A+B, 12 (3.2%) were positive by rRT-PCR. Compared with rRT-PCR, Sofia had a sensitivity of 100% (12/12; 95% CI: 73.5–100.0%) and a specificity of 61.2% (219/358; 95% CI: 55.9–66.3%). One-hundred thirty nine (38.8%) were false positive on Sofia. Of the 139 false positives, 123 (88.5%) were influenza B and 16 were influenza A. There was no difference in the time between illness onset and specimen collection date between true and false positives (median days = 2;
We learned that the assay lots we purchased were made after the recognized problem that led to the recall was corrected. Nevertheless, our data suggest that continuing problems exist. We share these data to alert others using this assay to the risk of false positives.
Our findings further support the limitations in using RIDTs for clinical decision making.
The findings and conclusions in this report are those of the authors and do not necessarily represent the views of the Centers for Disease Control and Prevention. The views and opinions expressed herein are the private opinions of the authors and do not reflect necessarily those of the US Army or the Department of Defense. The authors have no funding or conlflicts of interest to disclose.