Effectiveness of practices to reduce blood sample hemolysis in EDs: A laboratory medicine best practices systematic review and meta-analysis
-
Sep 2012
Source: Clin Biochem. 2012; 45(0):1012-1032. -
Alternative Title:Clin Biochem
-
Personal Author:
-
Description:Objective
To complete a systematic review of emergency department (ED) practices for reducing hemolysis in blood samples sent to the clinical laboratory for testing.
Results
A total of 16 studies met the review inclusion criteria (12 published and 4 unpublished). All 11 studies comparing new straight needle venipuncture with IV starts found a reduction in hemolysis rates, [average risk ratio of 0.16 (95% CI=0.11–0.24)]. Four studies on the effect of venipuncture location showed reduced hemolysis rates for the antecubital site [average risk ratio of 0.45 (95% CI=0.35–0.57].
Conclusions
Use of new straight needle venipuncture instead of IV starts is effective at reducing hemolysis rates in EDs, and is recommended as an evidence-based best practice. The overall strength of evidence rating is high and the effect size is substantial. Unpublished studies made an important contribution to the body of evidence. When IV starts must be used, observed rates of hemolysis may be substantially reduced by placing the IV at the antecubital site.
Disclaimer
The findings and conclusions in this article are those of the authors and do not necessarily represent the official position of the CDC.
-
Subjects:
-
Pubmed ID:22968086
-
Pubmed Central ID:PMC4518461
-
Document Type:
-
Funding:
-
Collection(s):
-
Main Document Checksum:
-
Download URL:
-
File Type:
-
jpeg gif jpeg gif jpeg gif jpeg bin gif jpeg gif jpeg gif jpeg gif
Details:
Supporting Files
More +