Electrophysiological Changes Correlated with Temperature Increases Induced by High-Intensity Focused Ultrasound Ablation
-
Dec 13 2014
Source: Ultrasound Med Biol. 2014; 41(2):432-448. -
Alternative Title:Ultrasound Med Biol
-
Personal Author:
-
Description:To gain better understanding of the detailed mechanisms of high-intensity focused ultrasound (HIFU) ablation for cardiac arrhythmias, we investigated how the cellular electrophysiological (EP) changes were correlated with temperature increases and thermal dose (cumulative equivalent minutes [CEM43]) during HIFU application using Langendorff-perfused rabbit hearts. Employing voltage-sensitive dye di-4-ANEPPS, we measured the EP and temperature during HIFU using simultaneous optical mapping and infrared imaging. Both action potential amplitude (APA) and action potential duration at 50% repolarization (APD50) decreased with temperature increases, and APD50 was more thermally sensitive than APA. EP and tissue changes were irreversible when HIFU-induced temperature increased above 52.3 ± 1.4°C and log10(CEM43) above 2.16 ± 0.51 (n = 5), but were reversible when temperature was below 50.1 ± 0.8°C and log10(CEM43) below -0.9 ± 0.3 (n = 9). EP and temperature/thermal dose changes were spatially correlated with HIFU-induced tissue necrosis surrounded by a transition zone.
-
Subjects:
-
Pubmed ID:25516446
-
Pubmed Central ID:PMC4297512
-